3.3.7 The calibration parameters
For each wavelength, the current calibration parameter is displayed as the new
parameter calculated. This will allow you to assess the new calibration in the
comparison. Too much difference between these values may be a sign of a
problem for your photometer (clogged or dirty sensor).
The three decimal correlation coefficient used to assess
the consistency of the measurements for calibration. A
good calibration should have a ratio above 0.99,
absolutely 1.
Updating coefficients
The "Update device" button to transmit the USB link the
new coefficients connected to the photometer.
Calibration bulletin publication
Once satisfied with your work, you must create the
calibration bulletin will file two forms: a text file and a
PDF file that are described in the following paragraph..
WARNING ! CREATE BULLETIN CALIBRATION DOES NOT AUTOMATICALLY
UPDATE THE FACTORS CALIBRATION PHOTOMETER !
3.3.8 Organization on the hard disk
For the photometer identified No xxxx, when the user chooses to produce a
calibration report, data is written to the file :
<username>
/CalitooData
/
xxxx/xxxx_CALIBRATION/xxxx_LANGLEY/
A text file and a PDF file is produced.