![Tenum CALITOO User Manual Download Page 33](http://html1.mh-extra.com/html/tenum/calitoo/calitoo_user-manual_1087852033.webp)
October 2016
33
The window consists of three areas:
- The Data Curves area displays the measurement points and the approximation
straight lines. You can enlarge them for better reading by double clicking it. The
same action makes them resume their original size. .
- The Parameters panel displays the numerical values of the initial calibration
coefficients and new coefficients calculated in real time and correlation
coefficients ( Deviation between the measurements curve and the straight line of
approximation. The closer it is to 1 , is the best approximation ) .
The Photometer update allows :
- update directly in the connected photometer, the three coefficients CN
0
with new values.
- create a calibration report in PDF format (that is the one that comes
printed in the box of your instrument) .
- The Data Source panel allows you to retrieve data according to the day, month
and year to which they were taken. We start by detailing the area which is the first
task to proceed calibration .
3.3.5 Data Source
Data are displayed as a calendar showing the years, months and days.
Day for which measures have been taken are displayed as red bars on the yellow
background of the month. The calendar starts with the photometer manufacture
date and ends the month of the current date.
Here the photometer was manufactured in November 2013 and measurements at
the beginning of February 2014.
Clicking the yellow rectangle in the month of February 2014 refines research
data :
We see that the available data correspond to measurements made on February
4
th
and 5
th
2014.
We will click on February 4
th
to see these measurements :
Summary of Contents for CALITOO
Page 49: ...October 2016 49 4 Appendix...