![Tenum CALITOO User Manual Download Page 38](http://html1.mh-extra.com/html/tenum/calitoo/calitoo_user-manual_1087852038.webp)
38
October 2016
3.3.9 Log file
At each change of calibration parameters, the software updates a log file. It is
located in the folder :
<username>
/CalitooData
/
xxxx/xxxx_Log.txt
for our example and is called :
0030_Log.txt
Here are its contents:
Calitoo #1302-0030
--------------
2013-12-14 13:14:32
---------------------------------------------------------------
CN0_465=3423;RAY_465=0.19490
CN0_540=3132;RAY_540=0.10637;OZ_540=0.0128
CN0_619=2455;RAY_619=0.06281;OZ_619=0.0154
2014-02-21 09:43:10
---------------------------------------------------------------
CN0_465=3582;RAY_465=0.19490
CN0_540=3154;RAY_540=0.10637;OZ_540=0.0128
CN0_619=2450;RAY_619=0.06281;OZ_619=0.0154
We see that the photometer No. 30 received the first calibration coefficients
December 14
th
, 2013 and new coefficients, February 21
th
, 2014.
This log will track the evolution of each meter and will help to refine the accuracy
of measurements.
Summary of Contents for CALITOO
Page 49: ...October 2016 49 4 Appendix...