43
•
To set the desired parameter, press the
corresponding MHz, then enter in the desired
value and select the unit as required.
Enabling Channel Output
•
Press the
CH1
button on the front panel to
switch on CH1 output directly, or press
Utility
after the label pops up, press
CH1 Setting
to
enable the output.
•
With CH1 output enabled, the CH1 button
backlight turns on and “Off” in grey to the right
of CH1 label changes into “Sweep” in yellow,
indicating CH1 output has been enabled.
•
Sweep waveform with an oscilloscope:
OUTPUTTING IMPULSE TRAIN
•
The Generator can create a cycle-specified waveform (name Impulse Train).
Internal, manual and external triggers are available to output the train.
•
There are three types of impulse trains: N-Cycle, Gated, and Infinite. The impulse
train can be produced with sine, square, ramp, pulse, arbitrary (except DC) or
noise (applicable for Gated type only) waveforms. CH1 and CH2 of the Generator
can be modulated independently with modulation types applied with the same one
or different ones.
Selecting Impulse Train
•
Enable impulse train function by pressing
Menu→Impulse Train
.
•
The Generator will output the pulse train according to the current settings.
Selecting Waveforms
•
N-Cycle Mode: Support sine, square, ramp, pulse and arbitrary (except DC)
waves.
•
Gated Mode: Support sine, square, ramp, pulse and arbitrary (except DC) waves
and noise.
•
Infinite Mode: Support sine, square, ramp, pulse and arbitrary (except DC) waves.