- 9 -
RS232 INTERFACE
For 10.1 should be open in setting menu.
Example of receipt like to below;
SETTINGS
o
Pres
F
key during 3 seconds to enter settings menu.
“Set” and “Basis Menu Number” is seen on Weight display, and “Sub Menu Number” is seen on Unit
Price display. Value is seen on Price to Pay display.
o
Press
F
key to pass to following menu.
o
Press
O
key to pass to following sub menu.
To change the type of
On
/
Off
values:
Press
T
key.
To change the type of number values:
Number keys should be used to change
Press
F
key to save the number.
o
Press
F
key for 3 seconds to return normal weighing.
NOTE: After exiting from Setting menu, scale should be turned off and turned on for a reliable weighing.
09.01.13 09:45 FİŞ NO: 000030
AĞIRLIK
:2.000
BİRİM AĞ.
:20 g
ADET
:100
- 10 -
EXAMPLE:
Turn off “Set 2-4” to block Automatic Zero Setting specification continuously. (2:basis menu
number, 4:sub menu number):
Press
F
key during 3 seconds.
“
Set1
” is seen on Weight display, 1 (submenu number) is seen on Unit Price display. Value is seen
on Price to Pay display.
Press
F
key once. Pass to menu 2. (Zero setting)
“
Set2
” is seen on Weight display, 1 (sub menu number) is seen on Unit Price display. Value is seen
on Price to Pay display.
Press
O
key 3 times. Pass to sub menu 4. (Automatic Zero setting)
“
Set2
” is seen on Weight display, “
4
” (sub menu number) is seen on Unit Price display. Value is
seen on Price to Pay display.
Press
T
key to make “
Off
” the value seen on Price to Pay display.
“
Set2
” is seen on Weight display, “
4
” (sub menu number) is seen on Unit Price display.
“
Off
” is
seen on Price to Pay display.
Press
F
key 3 seconds to return normal weighing.
Turn off the scale and turn it on again.
SETTINGS MENU:
(Values in parenthesis are factory settings.)
1. Tare
1.1 Tare depends on stable equilibrium
(On)
(If it is approved, its default value is ON and it
cannot be changed)
1.2 Cancel tare operation in negative net
value (Off)
Smaller tare than valid tare. (In negative net
value)
(If it is approved, its default value is ON and it
cannot be changed)
1.3 Tare from interface
(Off)
Not in this type.
2. Zero setting:
2.1 Initial zero setting range control ±10%
(Off)
(If it is approved, its default value is ON and it
cannot be changed)
2.2 Zero setting key range control < 2%
(Off)
(If it is approved, its default value is ON and it
cannot be changed)
2.3 Zero setting depends on stability (On)
(If it is approved, its default value is ON and it
cannot be changed)
2.4 Automatic zero setting
(Off)
If the weight indication is negative for more
than 5 seconds when the equilibrium indicator
is on, zero will be set automatically.
2.5 Automatic zero tracking
(Off)
If it is ON
: Automatic zero tracking is done
when the display is zero and stable.
If it is OFF : Automatic zero tracking will not
be done.
NOTE
: If there is tare, automatic zero
tracking program is not going to run.
2.6 Zero from interface
(Off)
Not in this type.
3. Total Memory, Unit Price:
3.1 Adding without showing total memory
(On)
If it is ON: Total cannot be seen after M+.
Weighing is seen.
If it is OFF: Total can be seen until the weighing
plate is empty, after M+
3.2 Temporary unit price
(Off)
Not in this type.
3.3 Adding by M+ is depends on stability
(On)
(If it is approved, its default value is ON and it
cannot be changed.)
4. Limitation:
Not
in
this
type.