2.3.
CONTROL PANEL (DISPLAY)
For the replacement of the Control panel (display), these steps should be followed:
a) Unplug the cord plug of the cabinet.
b) Remove the two M5 screws that hold the frontal cover and lift it up, fixing it with the two
rods provided for that objective (Fig.1)
c) Disconnect the connector of the display card (Fig.2)
d) Unscrew the six M3 screws and its washers (Fig.3)
e) Take out carefully the display card.
For the installation of the new display, all steps should be carried out in reverse order,
reminding to screw the two M5 screws that fix the frontal cover.
Fig. 1
Fig. 2
Fig. 3