SM-BioIIAdvancePlus-EN-1812
8 / 40
2.2.
CPU CARD
For the replacement of the CPU card, these steps should be followed:
a) Unplug the cord plug of the cabinet.
b) Remove the two M5 screws that hold the frontal cover and lift it up, fixing it with the two
rods provided for that objective. (Fig.1)
c) Disconnect all CPU card connectors (Fig.2)
d) Unscrew the M4 central screw and its washer (Fig.3)
e) Remove the four fixing screws and take out carefully the CPU card (Fig.4)
For the installation of the new CPU card, all steps should be carried out in reverse order,
reminding to connect the central grounding with its washer (Fig.3), and screwing the two M5
screws that fix the frontal cover. By replacing the chip (Fig.5) of the new
c
ard by the old one
the calibration data are maintained. Otherwise it is necessary to re-calibrate the cabinet.
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5