3.1.
REPLACEMENT OF THE EXTRACTION FILTER
WARNING
Before proceeding to replace the filters, ensure that the fans are fully stopped.
For the replacement of the extraction filter, these steps should be followed:
a)
Remove the two M5 screws that hold the frontal cover and lift it up, fixing it with the two
rods provided for that objective (Fig.1)
b)
Remove all M5 screws that hold the frontal cover of the plenum (Fig.2) and remove it.
The plenum can then be accessed, where the filters are located.
c)
Operate the two clamps (left and right) that hold the extraction filter (Figs.3&4), by
pushing or pulling the red levers they have.
d)
Take out the filter by the frontal part (Fig.5).
For the installation of the new filter, all steps should be carried out in reverse order, checking
that it is perfectly supported all along the upper and the lower gaskets. As last step it is
necessary to re-calibrate the cabinet.
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5