Page 20 of 20
TVP Manual (full) March 2007 CE rev 03.doc
2.01 Unpacking and inspection
CAUTION
Before unpacking the unit please verify that the packaging is in good condition, including the
tip & tell and shock watch devices. Check these carefully before accepting the unit. If they
have been activated contact the shipper and your supplier immediately.
The case is reusable and recyclable. Please treat the environment with respect. When
removing the unit from the case ensure that the zip loc bag containing the refrigeration line
connections, tool roll and other essential components are not discarded. The TVP is mounted
to the pallet with four packing bolts. The wooden pallet on which the unit is shipped must be
removed prior to installation.
Inspect the unit for any signs of damage during shipping. If damage is noted contact the
carrier immediately. For any further information contact Telemark or your distributor.
System shipping pressure
The TVP was shipped with an approximate standing pressure of 250psi the exact pressure,
which the unit shipped with, is noted on the packing list. A pressure gauge is mounted on the
green service panel on the lower right hand side of the unit. The static pressure may vary by
up to +/- 16 psi with ambient temperature.
Additional items shipped with the TVP
The unit has been shipped with a comprehensive set of tools and components to enable you to
install and maintain the unit. Please check all items listed on the packing list are present.
Replacement parts list including tools are included in section 4.
2.02 Operating environment
Before installation ensure that the following minimum services are available.
Service
TVP2000
TVP3500
Dimensions (L x W x H) mm
935 x 603 x 1745 mm
935 x 603 x 1745 mm
Floor space required (L x W)
1335 x 803 mm
1335 x 803 mm
Electrical Supply (400 volt compressor code TWD)
With recommended protection circuit breaker
value
380–420VAC/3-ph/50 Hz(35 Amps)
460 VAC/3-ph /60 Hz (35 Amps)
380– 420 VAC/3-ph/50 Hz
(60 Amps)
460 VAC/3-ph /60 Hz (60 Amps)
Electrical Supply (200 volt compressor code TWC)
With recommended protection circuit breaker
value
200 VAC/3-ph/50 Hz (60 Amps)
208–230 VAC/3-ph/60 Hz(60amps)
200 VAC/3-ph/50 Hz (90 Amps)
208–230 VAC/3-ph/60 HZ
(90Amps)
Water flow requirements at 15
o
c inlet temp (40
o
c
Outlet)
6 L / min
12 L / min
Heat dissipated to water at peak load kW
11
20.4
Shipping weight
384 kg
494 kg
Water connections
¾” NPS female
¾” NPS female
Noise level dba @ 1m
72
74.5
Refrigeration line connections Parker
2 x female Ultraseal supplied
2 x female Ultraseal supplied
Refrigeration line connections VCR
(optional)
VCR 8 male & female supplied
VCR 8 male & female supplied
Water quality
Ph 6.5 – 8 0.1 M ohm resisting
chloride free
Ph 6.5 – 8 0.1 M ohm resisting
chloride free