PS200
PORTABLE GAS MONITOR
MANUAL BUMP TEST OPTIONS
64171, Rev. 11/December 2019
6-4
6.
4.
1. BUMP TEST – PASS
After selecting
YES
, BUMP TEST PASS is displayed, as shown in
.
Figure
6-6: Bump Test Pass
The bump test result including date and time is automatically datalogged.
When the gas readings fall below their alarm set-points, or after 60 seconds, the bump test is com
-
plete, and the monitor will automatically return to normal operation.
6.
4
.2. BUMP TEST – FAIL
If any range fails the bump test, the CONFIRM ALARMS display will be red and a symbol will be
displayed, as shown in
Figure 6-7: Confirm Alarms - Fail
After selecting
YES
or
NO
, BUMP TEST FAIL is displayed and the user is instructed to switch the
Bump Test Fail - Switch Monitor OFF
.
Figure 6-8: Bump Test Fail - Switch Monitor OFF
If a monitor fails a bump test, then a full re-calibration is necessary.