PS200
PORTABLE GAS MONITOR
WARNINGS & FAULTS
64171, Rev. 11/December 2019
5-3
5.
3
.2. ZERO FAULT – DURING OPERATION
If ZERO FAULT is displayed during instrument operation, then:
• Leave the monitor ON for 30-60 minutes
• Switch OFF then ON
• If ZERO FAULT persists, recalibrate the monitor
• If ZERO FAULT still persists, return the monitor to an approved service/repair facility
5.
4
. Sensor Faults
If a sensor fault is detected, the backlight illuminates red, audible/visual alarms are activated, and a
wrench symbol is shown next to the faulty sensor.
5.
4
.1. SENSOR FAULT - LEL OR O
2
If a sensor fault is detected for LEL or O
2
, as illustrated in
, return
the monitor to an approved service/repair facility.
Figure
5-6: LEL SENSOR FAULT
5.
4
.2. SENSOR FAULT - CO OR H
2
S
If a sensor fault is detected for CO or H
2
S, as illustrated in
,
allow the monitor to run in clean air for 20 minutes. If the fault persists, return the monitor to an
approved service/repair facility.
Figure
5-7: CO/H
2
S SENSOR FAULT
5.
5
. Flow Fault (Pumped Monitors Only)
If a sample fault exists, the monitor displays a FLOW FAULT warning, as shown in
. The display will be red and both the audible alarm and red LED’s will be activated.