PS200
PORTABLE GAS MONITOR
OPERATION
64171, Rev. 11/December 2019
3-9
3.8. Sample Line Connection and Use
The monitor has an internal pump for remote sensing. Connect the sample line tubing to the sam
-
ple line connector, as shown in
. The pump is OFF after
start-up.
Figure 3-16: Sample Line Connection
3.8.1. PUMP OPERATION
Press and hold the R button to start or stop the pump.
When the pump is running, a pump symbol (as highlighted in
rotates in the display.
Figure 3-17: Pump Symbol
3.9. Self Test
The monitor can perform a self test at any time during normal operation.
To perform a self test, press and hold the L button.
The monitor tests the buzzer, the LED’s, and the vibration function. The flammable gas used for
calibration is displayed, as illustrated in
NOTE: It is only possible to switch the pump ON / OFF when monitor alarms are inactive.
NOTE: Pump cannot be switched OFF if monitor is configured with setting ‘PUMP
ALWAYS ON’.