Summary of Contents for CP030

Page 1: ...Operator s Manual CP030 CP030A CP031 CP031A Current Probes ...

Page 2: ......

Page 3: ...CP03x CP03xA Current Probes Operator s Manual September 2016 ...

Page 4: ...are encouraged to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes Teledyne LeCroy is a registered trademark of Teledyne LeCroy Inc Windows is a registered trademark of Microsoft Corporation Other product or brand names are trademarks or requested trademarks of their respective holders Information in this publication supersedes all earlier versions Specifica...

Page 5: ...e Maximum Input Current vs Frequency 7 CP030 CP030A Insertion Impedance vs Frequency 7 CP031 CP031A Derating Curve Maximum Input Current vs Frequency 8 CP031 CP031A Insertion Impedance vs Frequency 8 Operation 9 Precautions 9 Connecting to the Test Instrument 10 Connecting to the Test Circuit 10 Operating with an Oscilloscope 11 Performance Verification 13 Test Equipment Required 14 Preliminary Pr...

Page 6: ...nd Maintenance 18 Cleaning 18 Calibration Interval 18 Service Strategy 18 Troubleshooting 18 Returning a Product for Service 19 Teledyne LeCroy Service Centers 20 Warranty 21 Certifications 22 EMC Compliance 22 Safety Compliance 23 Environmental Compliance 24 ...

Page 7: ...y or damage Do not proceed until conditions are fully understood and met Do not apply around or remove from UNINSULATED HAZARDOUS LIVE conductors which may render electric shock electric burn or arc flash Equipment protected by double insulation or reinforced insulation CAUTION hot surface Precautions Comply with the following safety precautions to avoid personal injury or damage to your equipment...

Page 8: ...the insulation could cause an electric shock Use only indoors and within the operational environment listed Do not use in wet or explosive atmospheres Do not remove the probe s casing Touching exposed connections may result in electric shock Keep product surfaces clean and dry Comply with the maximum input current vs frequency derating when measuring current that includes a high frequency componen...

Page 9: ...sional or later The CP03xA probes are compatible with most Teledyne LeCroy oscilloscopes with a ProBus interface running X Stream firmware version 7 8 x x or later See table below Compatible Compatible with Upgrade Not Compatible LabMaster 9MZi A DDA SDA WaveMaster 8 Zi Zi A Zi B DDA SDA WavePro 7 Zi Zi A HRO WaveRunner 6 Zi WaveRunner Xi Xi A Mxi Mxi A HDO8000 HDO6000 HDO4000 MDA800 WaveSurfer 10...

Page 10: ...10 MHz DC to 100 MHz Nominal Characteristics CP030 CP030A CP031 CP031A Max Continuous Input Current at 0 1 V A 30 Arms 30 Arms Max Peak Current 50 Apeak non continuous 50 Apeak non continuous Intended Output Load 1 MΩ 1 MΩ Measured Conductors Insulated Conductors only Insulated conductors only Sensitivity 10 mA div to 15 A div 1 mA div to 15 A div 10 mA div to 15 A div 1 mA div to 15 A div Offset ...

Page 11: ...031A Operating Temperature 0 C to 40 C 0 C to 40 C Operating RH 80 non condensing 80 non condensing Storage Temperature 10 C to 50 C 10 C to 50 C Storage RH 80 non condensing 80 non condensing Maximum Altitude 2000 m 2000 m Effect of External Magnetic Field Maximum of 20 mA in a DC 60 Hz 400 A m magnetic field Maximum of 5 mA in a DC 60 Hz 400 A m magnetic field Physical Characteristics CP030 CP03...

Page 12: ...or measurements within Measurement Categories II III or IV 3 Not rated for measurements on uninsulated conductors 4 Pollution Degree 2 refers to operating environment where normally only dry non conductive pollution occurs Temporary conductivity caused by condensation should be expected NOTE Because of the marking requirements per safety standard EN 61010 2 032 2012 Third Edition the measurement c...

Page 13: ...Operator s Manual 7 CP030 CP030A Derating Curve Maximum Input Current vs Frequency Does not apply to CP030 3M CP030 CP030A Insertion Impedance vs Frequency Does not apply to CP030 3M ...

Page 14: ...CP03x CP03xA Current Probes 8 CP031 CP031A Derating Curve Maximum Input Current vs Frequency CP031 CP031A Insertion Impedance vs Frequency ...

Page 15: ...to strong external electric fields WARNING Comply with the derating characteristics of the probe when measuring current that includes a high frequency component and never measure any current that exceeds the rated current Using the probe at high frequencies or in strong magnetic fields may cause the probe head to become abnormally hot resulting in fire equipment damage or burns CAUTION Do not touc...

Page 16: ...ned with a movable split core eliminating the need to break the conductor for the core to slip around the conductor To connect 1 Pull back the slider so that the clamp opens 2 Align the sensor so that the current direction indicator corresponds to the direction of current flow in the conductor 3 Close the slider on the sensor head until the UNLOCK indication disappears and the LOCK indication appe...

Page 17: ...e is detected The Probe dialog also contains controls for Auto Zero and Degauss Probe dialog Probe A div and Attenuation The front panel Volts div knob controls the oscilloscope s scale factor and the probe s internal attenuation to give full available dynamic range from 10mA div to 15 A div for CP030 and CP031 or 1 mA div to 15 A div for CP030A and CP031A CP031A and CP031A will automatically adju...

Page 18: ...t is important not to exceed the maximum linear input values when using AC coupling The maximum DC input current is 30A when the probe is in 0 1 V A sensitivity 100mA div and 5A when the probe is in 1 V A 100mA div Auto Zero The Probe dialog incorporates an Auto Zero function to remove the DC offset from the current probe Auto Zero must be invoked by the user After several minutes of warm up or wh...

Page 19: ...of the Test Record provided at the end of the manual The warranted characteristics of the probe are valid at any temperature within the Environmental Characteristics listed in the Specifications However some of the other test equipment used to verify the performance may have environmental limitations required to meet the accuracy needed for the procedure Make sure that the ambient conditions meet ...

Page 20: ...Stanford Research Model DS340 Calibration Fixture 100 turn loop in series with 0 5Ω 0 1 resistor with sense terminals Teledyne LeCroy CP031 CF02 Calibration Fixture ProBus Extension Cable Teledyne LeCroy PROBUS CF01 Banana Plug Adapter Female BNC to Dual Banana Plug Pomona 1269 Patch Cables 4 Male Banana to Male Banana 12 Pomona B 12 0 black Pomona B 12 2 red BNC Adapter BNC Male to Dual Banana Ja...

Page 21: ... C1 setup dialog and confirm that the bandwidth is set to Full BWL OFF 2 Verify that the probe is sensed and the probe dialog CP03x or CP03xA tab appears behind the C1 setup dialog 3 Open the probe dialog and degauss the probe by touching DEGAUSS then OK 4 Confirm that the message Performing Degauss on CP03x is displayed in the message bar and that no error messages are displayed Check LF Accuracy...

Page 22: ...DEGAUSS button 7 Set the current probe channel sensitivity to 1 A div 8 Open the probe slider and position the probe input around the 100 Turn loop Close and LOCK the slider 9 Adjust the Function generator voltage until the voltage measured at the Current Sense terminals DMM 2 reads 50 mV 0 05 mV This corresponds to 10 A at the probe head 10 Record the voltage measured by DMM 1 on the Test Record ...

Page 23: ..._________________________________ Equipment Model Serial Number Calibration Due Date Digital Multimeter 1 Digital Multimeter 2 Function Generator N A The function generator is used for making relative measurements The output of the generator is measured with a DMM or oscilloscope Thus the generator is not required to be calibrated Step Description Intermediate Data Test Result 10 Probe Output limi...

Page 24: ...ility for diagnosis and exchange A defective probe under warranty will be replaced with a factory refurbished probe A probe that is not under warranty can be exchanged for a factory refurbished probe A modest fee is charged for this service A defective probe must be returned in order to receive credit for the probe core Troubleshooting If the probe is not operating properly the problem may be the ...

Page 25: ... material or equivalent 3 Label the case with a tag containing The RMA Name and address of the owner Probe model and serial number Description of failure or requisite service 4 Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit 5 Mark the outside of the box with the shipping address given to you by Teledyne LeCroy be sure to add the following ATTN R...

Page 26: ... 3411217 Meyrin 1 Geneva Switzerland teledynelecroy com Europe Ph 41 22 719 2111 FAX 41 22 719 2230 contact sa teledynelecroy com China LeCroy Corporation Beijing Rm 2001 Unit A Horizon Plaza No 6 Zhichun Road Haidian Dist Beijing 100088 China www lecroy com cn Sales Ph 86 10 82800318 0319 0320 FAX 86 10 82800316 Marketing China teledynelecroy com Service Rm 2002 Ph 86 10 82800245 Service China te...

Page 27: ...efective due to workmanship or materials and the defect is not caused by misuse neglect accident abnormal conditions of operation or damage resulting from attempted repair or modifications by a non authorized service facility The customer will be responsible for the transportation and insurance charges for the return of products to the service facility Teledyne LeCroy will return all products unde...

Page 28: ...adiated and Conducted Emissions Group 1 Class A 2 3 ELECTROMAGNETIC IMMUNITY EN 61000 4 2 2009 Electrostatic Discharge 4 kV contact 8 kV air 4 kV vertical horizontal coupling planes 4 EN 61000 4 3 A2 2010 RF Radiated Electromagnetic Field 3 V m 80 1000 MHz 3 V m 1400 MHz 2 GHz 1 V m 2 GHz 2 7 GHz EN 61000 4 8 2010 Power Frequency Magnetic Field 3 A m 50 Hz 3 A m 60 Hz 1 To ensure compliance with t...

Page 29: ...S Components Ltd Unit 30 31 Warehouse World 761 Great South Road Penrose Auckland New Zealand Visit teledynelecroy com support contact for the latest contact information Safety Compliance EC Declaration of Conformity Low Voltage The probe meets the intent of EC Directive 2014 35 EU for Product Safety Compliance was demonstrated to the following specifications as listed in the Official Journal of t...

Page 30: ...tronic Equipment WEEE and Batteries The probe is subject to disposal and recycling regulations that vary by country and region Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles For more information about proper disposal and recycling of your Teledyne LeCroy product visit teledynelecroy com recycle Restriction of Hazardous Substances RoHS The product a...