Operator’s Manual
15
Functional Check
The functional check will verify the basic operation of the probe functions. It is
recommended to perform the functional check prior to the performance verification
procedure.
1.
Open the C1 setup dialog and confirm that the bandwidth is set to
Full (
BWL
OFF).
2.
Verify that the probe is sensed and the probe dialog (
CP03x or CP03xA tab
)
appears behind the C1 setup dialog.
3.
Open the probe dialog and degauss the probe by touching
DEGAUSS
, then
OK
.
4.
Confirm that the message
"Performing Degauss on CP03x...."
is displayed in the
message bar and that no error messages are displayed.
Check LF Accuracy
LF Accuracy Test set up.