CP03x/CP03xA Current Probes
6
Safety Ratings
Probe Designation
1
Type D
Measurement Category
2
No rated measurement category
Safe Voltage Rating
3
Use only on fully insulated conductors
Pollution Degree
4
2
Definitions (per IEC/EN 61010-2-032:2012)
1.
Type D current probe is intended to be applied around or removed from insulated conductors.
2.
No rated measurement category; not intended for measurements on circuits directly connected to the
Mains supply. Not rated for measurements within Measurement Categories II, III, or IV.
3.
Not rated for measurements on uninsulated conductors.
4.
Pollution Degree 2 refers to operating environment where normally only dry, non-conductive pollution
occurs. Temporary conductivity caused by condensation should be expected.
NOTE:
Because of the marking requirements per safety standard EN 61010-2-032:2012 (Third Edition), the
measurement category rating that was on the CPxxx probes originally certified per EN 61010-2-032:2002
(Second Edition) has been removed from the product label. However, this change does not reduce the
measurement capability or the level of protection offered by the new probe design.