background image

page 9

back to atmospheric settings. A long press of the pressure correct button changes the 
unit back to the Viewing Dewpoint Mode. Pressing the ‘Mode’ button changes the unit to 
the View/Set Gas Pressure sub-state. The display has the ‘SET’ and ‘PSI’ legends on 
and alternately shows 

*$6

and the currently set value for the gas pressure. The up, down 

and pressure correct buttons operate in the same manner as in the Sensor Pressure 
sub-state. Pressing the ‘Mode’ button changes the unit back to View/Set Sensor Pres-
sure sub-state, and so forth.
The factory default settings are: 14.7psi for both sensor and gas pressure and pressure 
correction disabled.
When Pressure correction is disabled all dewpoints are computed by assuming that both 
Sensor and Gas Pressures are 14.7psi.

5.2.2 Viewing Temperature at Sensor

In this mode the user can view the temperature at the sensor inside the sampling cham-
ber, signified by the flashing ‘TEMP’ legend on the lower left of the display. The available 
units in which to view the temperature are °C and °F. The up and down buttons switch 
back and forth between these units.
Pressing the ‘Mode’ button changes to the Start Autocal Mode.

5.2.3 Start Autocal Mode

In this mode the user is prompted to confirm that Autocal should be initiated. This mode 
can be skipped by pressing the ‘Mode’ button to go to the View Serial Number Mode or 
by cancelling with the down or pressure correct buttons which displays the word 

&$1

 for 

1/2 second and goes back to the Viewing Dewpoint Mode.
Autocal can be initiated by pressing the ‘up’ button. The unit will check if the sensor is in 
the sampling chamber, if not it will prompt the user to pull the sensor out of the desiccant 
by beeping and displaying alternately 

38/

 and 

287

. Pull the sensor actuator out, when 

this message appears.
Since Autocal is a irreversible recalibration of the instrument there is one more beeping 
and visual prompting with alternating 

&1)

 and

$&"

. The down and pressure correct but-

tons allow cancelling the Autocal, while the up button confirms it. The unit starts taking 
measurements for 1 minute while displaying alternately flashing 

$&

. During this time the 

user should occasionally pump saturated air using the calibration bulb into the sample 
chamber (see Autocal Instructions). An alternating 

6(1

 and 

/2

 after one minute, 

accompanied by a beeping, indicates the sensor has a reading too low to be from a satu-
rated sensor. This error will cause Autocal to be cancelled. However, if the sensor read-
ing is within range of a saturated sensor, a new calibration is calculated (some numbers 
flash on the display),

(1'

is displayed for 1 second, and the new calibration values are 

saved in EEPROM.

5.2.4 Viewing Serial Number Mode

In this mode the user can view the serial number of the sensor. The display shows the 
serial number by alternately displaying 

;61

 and the number. If the number is larger than 

1999 then it is displayed in 2 segments, first segment is the thousands signified by the 

x10

3

 legend in the upper right corner of the display and the second segment is the units. 

Summary of Contents for 8800P

Page 1: ...instruction manual Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies Model 8800P TRACE MOISTURE ANALYZER...

Page 2: ...presentative is Except as may be provided by contract this document and all specifica tions and drawings contained are the property of Teledyne Analytical Instruments are issued in strict confidence a...

Page 3: ...siccant cartridge and lithium battery 2 2 VCO to Swagelok fittings and 2 Swagelok to barbed hose fittings 3 Calibration bulb 4 Shoulder strap Optional Equipment 5 VCO to VCO Fittings 6 Pig tail 7 Flex...

Page 4: ...ressures slightly above atmospheric however each sample cell sensor piston dry storage assembly has been tested for pressures up to 120 PSI Do not introduce sample streams which could cause pressure r...

Page 5: ...mple Cell 10 Electronic Board 5 Sensor 11 Instrument Case 6 Spring loaded PTFE seals The drawing on the left shows the 8800P with the sensor in dry storage position The sen sor is enclosed in a gas ti...

Page 6: ...contamination with desiccant particles The factory supplied desiccant provides dry storage dewpoints as low as 80 C when fresh When the sensor is pushed into dry storage it will take some time to dry...

Page 7: ...tubes are perfectly acceptable and offer advantages due to their flexibility for dew points above 65 C However PTFE tubes should be kept as short as possible Under no circumstances use rubber or plas...

Page 8: ...ed if the dewpoint is decreasing Make sure the reading has completely stabilized before taking the final reading After the reading is taken push the sensor actuator back into the dry down position Not...

Page 9: ...f the display Be careful not to change any parameter inadvertently 3 C F appear simultaneously to indicate the sensors attenuation in decibels 4 Values higher than 1999 and lower than 0 01 are display...

Page 10: ...in and out of pressure cor rect mode When there is no pressure correction applied the PSI legend does not appear The PSI legend flashes at the bottom of the display when there is pressure cor rection...

Page 11: ...splays the word 1 for 1 2 second and goes back to the Viewing Dewpoint Mode Autocal can be initiated by pressing the up button The unit will check if the sensor is in the sampling chamber if not it wi...

Page 12: ...r dewpoint can be alternated in two different units 3 Selecting the sensor type In this mode the user can select the software matching the type of sensor installed in the instrument XTR 100 100 C to 2...

Page 13: ...ared see the pin out table below In order to receive the dewpoint in the currently selected engineering units the host com puter must query the 8800P with C R The 8800P will respond with the dewpoint...

Page 14: ...ull range of the sensor this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range or vise versa To verify or change the current loop scaling follow...

Page 15: ...VDC to 25VDC when operating from external DC power The unit requires 12VACrms to 25VACrms when operating from external AC power In either case the unit can draw up to 50mA when the backlight is on and...

Page 16: ...icro climate in sample cell while the display is flashing B 10 After one minute the instrument displays calibration numbers to be stored in the instruments memory displays 1 and thereafter goes back i...

Page 17: ...e sensor actuator The replacement cartridge is in a sealed tin container Open the container and remove the replacement desiccant cartridge Remove the GREEN cap DO NOT open the RED cap as the desiccant...

Page 18: ...to 50 C 231 Sensor circuit is open dewpoint displays tests check sensor wetup sensor to see if it will react 6 7 Sensor is saturated dry down the sensor 6 5 Sensor circuit is shorted check sensor Tryi...

Page 19: ...PRESS LONG PRESS SHORT PRESS 1 2 sec LONG PRESS AND PSI MODE 1 minute few sec 1 sec OUT OK TOO LOW IN DESICCANT 65 20 CAL XSN 123 END SEN 2LO CNF AC AC _AC PUL OUT CAN SEN VALUE GAS VALUE F C DEWPOINT...

Page 20: ...ve attenuation Manual calibration Set dewpoint corresponding to low end of analog output Factory default is low range of sensor e g 100 C for XTR 100 Factory default is high range of sensor e g 20 C f...

Page 21: ...4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 0 100 200 300 400 500 600 700 800 900 1000 1100 XTR100 4 20 mA output Instrument reading in ppmV...

Page 22: ...3 0 04 0 05 0 06 0 07 0 08 0 09 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 2 3 4 5 6 7 8 9 10 20 30 Measured Dewpoint C Dewpoint 3 C Dewpoint 3 C Uncertainty of ppmV calculation due to 3 C measurment accur...

Page 23: ...page 21 Appendix E Battery Life...

Reviews: