background image

page 12

If the current loop is not being used, it is safe to connect the DB9 connector directly to a 
PC; pins 1&9 will not be dammaged or couse dammage. 
The RS-232 levels are generated by the instrument only when an RS-232 level is 
detected at the DB9 input. Therefore, when operating from a battery to conserve its 
power, disconnect the RS-232 cable unless necessary for the operation.

6.2 Current Loop Output

The current loop signal is provided on the 9 pin sub ‘D’ female connector located at the 
rear of the instrument. The current loop output functions only when the instrument is 
operated from an external power supply. Pin 1 is the positive current loop output while pin 
5 is the ground.
This output is isolated from the sensor and case, however as a sefety precoution, there is 
a 1M

 shunt to prevent electrostatic buildup and discharge.

 Note that this connector is shared with the RS-232C interface, see section 6.1 for full pin 
assignments. Pin 9 controls the range of the current loop, when grounded the output is 4-
20mA, when left open the output is 0-24mA. The current loop is capable of driving loads 
from 0

Ω 

to 500

Ω. 

The output is linearly proportional to the selected engineering units. 

The output may be scaled such that it spans only a portion of the full range of the sensor, 
this feature may be useful in cases where a higher resolution output is required over a 
narrow dewpoint range, or vise versa. To verify or change the current loop scaling follow 
the instructions in the set-up mode section 5.3.

The current output is computed by the microprocessor as follows:

If pin 9 is grounded for a 4-20mA range:

Where:

O

=output current in mA.

R

= value of Reading shown on display in selected engineering units.

L

= value of Low Limit of Output Scaling converted to selected engineering units.

H

= value of High Limit of Output Scaling converted to selected engineering units.

See section 5.3 to select values for L & H. Their factory defaults are the upper and lower 
range of the sensor e.i. for XTR-100 L= -100°C and H= +20°C. The values for L & H are 
set up always in °C, however the microprocessor converts them to the currently selected 
units. Note that because R, L and H are all in the same units the output current is linearly 
proportional to those units. The current is linearly proportional to dewpoint, if the selected 

units are °C or °F, however if the selected units are Lbs, ppm or G/M

then the current is 

approximately logarithmically proportional to dewpoint see Appendix C.
After hooking up the current loop output, it can be forced to its low, mid and high points 
by following the instructions in section 5.3. This procedure may be helpful in testing the 
connection and setting-up the termination equipment.

O

R

L

(

)

H

L

(

)

------------------

16

4

+

×

=

Summary of Contents for 8800P

Page 1: ...instruction manual Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies Model 8800P TRACE MOISTURE ANALYZER...

Page 2: ...presentative is Except as may be provided by contract this document and all specifica tions and drawings contained are the property of Teledyne Analytical Instruments are issued in strict confidence a...

Page 3: ...siccant cartridge and lithium battery 2 2 VCO to Swagelok fittings and 2 Swagelok to barbed hose fittings 3 Calibration bulb 4 Shoulder strap Optional Equipment 5 VCO to VCO Fittings 6 Pig tail 7 Flex...

Page 4: ...ressures slightly above atmospheric however each sample cell sensor piston dry storage assembly has been tested for pressures up to 120 PSI Do not introduce sample streams which could cause pressure r...

Page 5: ...mple Cell 10 Electronic Board 5 Sensor 11 Instrument Case 6 Spring loaded PTFE seals The drawing on the left shows the 8800P with the sensor in dry storage position The sen sor is enclosed in a gas ti...

Page 6: ...contamination with desiccant particles The factory supplied desiccant provides dry storage dewpoints as low as 80 C when fresh When the sensor is pushed into dry storage it will take some time to dry...

Page 7: ...tubes are perfectly acceptable and offer advantages due to their flexibility for dew points above 65 C However PTFE tubes should be kept as short as possible Under no circumstances use rubber or plas...

Page 8: ...ed if the dewpoint is decreasing Make sure the reading has completely stabilized before taking the final reading After the reading is taken push the sensor actuator back into the dry down position Not...

Page 9: ...f the display Be careful not to change any parameter inadvertently 3 C F appear simultaneously to indicate the sensors attenuation in decibels 4 Values higher than 1999 and lower than 0 01 are display...

Page 10: ...in and out of pressure cor rect mode When there is no pressure correction applied the PSI legend does not appear The PSI legend flashes at the bottom of the display when there is pressure cor rection...

Page 11: ...splays the word 1 for 1 2 second and goes back to the Viewing Dewpoint Mode Autocal can be initiated by pressing the up button The unit will check if the sensor is in the sampling chamber if not it wi...

Page 12: ...r dewpoint can be alternated in two different units 3 Selecting the sensor type In this mode the user can select the software matching the type of sensor installed in the instrument XTR 100 100 C to 2...

Page 13: ...ared see the pin out table below In order to receive the dewpoint in the currently selected engineering units the host com puter must query the 8800P with C R The 8800P will respond with the dewpoint...

Page 14: ...ull range of the sensor this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range or vise versa To verify or change the current loop scaling follow...

Page 15: ...VDC to 25VDC when operating from external DC power The unit requires 12VACrms to 25VACrms when operating from external AC power In either case the unit can draw up to 50mA when the backlight is on and...

Page 16: ...icro climate in sample cell while the display is flashing B 10 After one minute the instrument displays calibration numbers to be stored in the instruments memory displays 1 and thereafter goes back i...

Page 17: ...e sensor actuator The replacement cartridge is in a sealed tin container Open the container and remove the replacement desiccant cartridge Remove the GREEN cap DO NOT open the RED cap as the desiccant...

Page 18: ...to 50 C 231 Sensor circuit is open dewpoint displays tests check sensor wetup sensor to see if it will react 6 7 Sensor is saturated dry down the sensor 6 5 Sensor circuit is shorted check sensor Tryi...

Page 19: ...PRESS LONG PRESS SHORT PRESS 1 2 sec LONG PRESS AND PSI MODE 1 minute few sec 1 sec OUT OK TOO LOW IN DESICCANT 65 20 CAL XSN 123 END SEN 2LO CNF AC AC _AC PUL OUT CAN SEN VALUE GAS VALUE F C DEWPOINT...

Page 20: ...ve attenuation Manual calibration Set dewpoint corresponding to low end of analog output Factory default is low range of sensor e g 100 C for XTR 100 Factory default is high range of sensor e g 20 C f...

Page 21: ...4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 0 100 200 300 400 500 600 700 800 900 1000 1100 XTR100 4 20 mA output Instrument reading in ppmV...

Page 22: ...3 0 04 0 05 0 06 0 07 0 08 0 09 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 2 3 4 5 6 7 8 9 10 20 30 Measured Dewpoint C Dewpoint 3 C Dewpoint 3 C Uncertainty of ppmV calculation due to 3 C measurment accur...

Page 23: ...page 21 Appendix E Battery Life...

Reviews: