background image

page 8

5.2 Operating State

Refer to Appendix A for a flow diagram of the operating state. The 8800P is powered ON 
by pressing the ‘Power’ button until a beep is heard; to activate the backlight, continue to 
depress the ‘Power’ button until the backlight turns on. Keep in mind that with the back-
light on the unit consumes about 5 times more power.
Upon power up the unit performs certain initialization tests (see section 9), and enters 
the ‘Operating State’, in the Viewing Dewpoint mode. Depressing the ‘Mode’ button will 
change modes in the following order: Viewing Dewpoint 

 Viewing Sensor Temperature 

 Start Autocal 

 Viewing Serial Number 

 (back to) Viewing Dewpoint. The unit will 

return to Viewing Dewpoint mode if no buttons are pressed for 30 seconds, unless it is 
performing Autocal.

5.2.1 Viewing Dewpoint Mode

In this mode the user can view the dewpoint, signified by the presence of the ‘DEW-
POINT’ legend on the lower left of the display. The available engineering units in which to 

view the moisture content are °C, °F, PPM, LBS and G/M

3

; the up and down buttons 

scroll back and forth through these units in respective order. The °C and °F are dewpoint 
readings. The ppm is parts per million by volume computed at the sensor pressure (more 

about pressure later). The Lbs and G/M

3

 are pounds of water per million standard cubic 

feet and grams of water per standard cubic meters, both in Natural Gas, they are com-
puted according to data derived by IGT Research Bulletin 8, taking into account sensor 
pressure. 
A short press of the ‘pressure correct’ button toggles the unit in and out of pressure cor-
rect mode. When there is no pressure correction applied the PSI legend does not 
appear. The PSI legend flashes at the bottom of the display, when there is pressure cor-
rection in the computation of the displayed values.
Pressure correction is used in the context that the values displayed signify the moisture 
content at some pressure (we refer to this as the ‘Gas Pressure’) different than the pres-

sure at the sensor. Note that PPM, LBS and G/M

3

 readings are by definition unaffected 

by pressure correction because only the pressure at the sensor affects their value. While 
°C and °F are affected by pressure correction by reporting what the dewpoint would be at 
the Gas Pressure when the dewpoint is what is measured at the pressure at the sensor. 
However, this also implies that whether pressure correction is applied or not the PPM, 

LBS and G/M

3

 readings are affected by the setting of the sensor pressure.

Sensor pressure is used in the context that this is the pressure inside the sampling 
chamber when performing the measurement.
Gas pressure is used in the context that this is the pressure to be used in computing 
what the dewpoint would be at a pressure different than the one in the sampling cham-
ber.
A long press of the Pressure Correct button, while in the pressure correct mode (flashing 
PSI legend), changes the unit to the View/Set Sensor Pressure sub-state. The display 
has the ‘SET’ and ‘PSI’ legends on and alternately shows 

6(1

and the currently set value 

for the sensor pressure. The up and down buttons allow the user to modify the sensor 
pressure, while a short press of the pressure correct button toggles the Sensor Pressure 
setting between whatever value is on the display and 14.7 psi ---- this is a quick way to go 

Summary of Contents for 8800P

Page 1: ...instruction manual Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies Model 8800P TRACE MOISTURE ANALYZER...

Page 2: ...presentative is Except as may be provided by contract this document and all specifica tions and drawings contained are the property of Teledyne Analytical Instruments are issued in strict confidence a...

Page 3: ...siccant cartridge and lithium battery 2 2 VCO to Swagelok fittings and 2 Swagelok to barbed hose fittings 3 Calibration bulb 4 Shoulder strap Optional Equipment 5 VCO to VCO Fittings 6 Pig tail 7 Flex...

Page 4: ...ressures slightly above atmospheric however each sample cell sensor piston dry storage assembly has been tested for pressures up to 120 PSI Do not introduce sample streams which could cause pressure r...

Page 5: ...mple Cell 10 Electronic Board 5 Sensor 11 Instrument Case 6 Spring loaded PTFE seals The drawing on the left shows the 8800P with the sensor in dry storage position The sen sor is enclosed in a gas ti...

Page 6: ...contamination with desiccant particles The factory supplied desiccant provides dry storage dewpoints as low as 80 C when fresh When the sensor is pushed into dry storage it will take some time to dry...

Page 7: ...tubes are perfectly acceptable and offer advantages due to their flexibility for dew points above 65 C However PTFE tubes should be kept as short as possible Under no circumstances use rubber or plas...

Page 8: ...ed if the dewpoint is decreasing Make sure the reading has completely stabilized before taking the final reading After the reading is taken push the sensor actuator back into the dry down position Not...

Page 9: ...f the display Be careful not to change any parameter inadvertently 3 C F appear simultaneously to indicate the sensors attenuation in decibels 4 Values higher than 1999 and lower than 0 01 are display...

Page 10: ...in and out of pressure cor rect mode When there is no pressure correction applied the PSI legend does not appear The PSI legend flashes at the bottom of the display when there is pressure cor rection...

Page 11: ...splays the word 1 for 1 2 second and goes back to the Viewing Dewpoint Mode Autocal can be initiated by pressing the up button The unit will check if the sensor is in the sampling chamber if not it wi...

Page 12: ...r dewpoint can be alternated in two different units 3 Selecting the sensor type In this mode the user can select the software matching the type of sensor installed in the instrument XTR 100 100 C to 2...

Page 13: ...ared see the pin out table below In order to receive the dewpoint in the currently selected engineering units the host com puter must query the 8800P with C R The 8800P will respond with the dewpoint...

Page 14: ...ull range of the sensor this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range or vise versa To verify or change the current loop scaling follow...

Page 15: ...VDC to 25VDC when operating from external DC power The unit requires 12VACrms to 25VACrms when operating from external AC power In either case the unit can draw up to 50mA when the backlight is on and...

Page 16: ...icro climate in sample cell while the display is flashing B 10 After one minute the instrument displays calibration numbers to be stored in the instruments memory displays 1 and thereafter goes back i...

Page 17: ...e sensor actuator The replacement cartridge is in a sealed tin container Open the container and remove the replacement desiccant cartridge Remove the GREEN cap DO NOT open the RED cap as the desiccant...

Page 18: ...to 50 C 231 Sensor circuit is open dewpoint displays tests check sensor wetup sensor to see if it will react 6 7 Sensor is saturated dry down the sensor 6 5 Sensor circuit is shorted check sensor Tryi...

Page 19: ...PRESS LONG PRESS SHORT PRESS 1 2 sec LONG PRESS AND PSI MODE 1 minute few sec 1 sec OUT OK TOO LOW IN DESICCANT 65 20 CAL XSN 123 END SEN 2LO CNF AC AC _AC PUL OUT CAN SEN VALUE GAS VALUE F C DEWPOINT...

Page 20: ...ve attenuation Manual calibration Set dewpoint corresponding to low end of analog output Factory default is low range of sensor e g 100 C for XTR 100 Factory default is high range of sensor e g 20 C f...

Page 21: ...4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 0 100 200 300 400 500 600 700 800 900 1000 1100 XTR100 4 20 mA output Instrument reading in ppmV...

Page 22: ...3 0 04 0 05 0 06 0 07 0 08 0 09 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 2 3 4 5 6 7 8 9 10 20 30 Measured Dewpoint C Dewpoint 3 C Dewpoint 3 C Uncertainty of ppmV calculation due to 3 C measurment accur...

Page 23: ...page 21 Appendix E Battery Life...

Reviews: