Oil in Water Analyzer
Operation 3
Part I 3-15
Teledyne Analytical Instruments
DIG_FILT SELF-TEST
PWD LOGOUT MORE
2.
Using the Right or Left arrow keys, select MORE and press
Enter. The second System screen appears:
AUTOCAL HMGNZR TRACK
CAL-HOLD-TIMER MORE
or
AUTOCAL HMGNZR HOLD
CAL-HOLD-TIMER MORE
3.
Using the Right or the Left arrow keys, select MORE and
press Enter. The third System screen appears:
ALGORITHM APPLICATION
MODEL OUT_CAL ANLZ
4.
Select OUTPUT_CAL and Enter
Use UP/DOWN arrow to
Adjust 4 ma: 250
The number on the second row is the setpoint of the 4 mA output. It is
analogous to a potentiometer wiper.
The number can be set anywhere from 0
to 500. The default is 250, in the middle. At the default setting, the output
should be very close to 4 mA. If not, slowly adjust the number using the Up or
the Down keys until DMM reads 4.00 mA. Enter when done.
5.
A screen similar to the one above will appear and the DMM
should read close to 20 mA. If not, slowly adjust the number using the Up or
Down key until DMM reads 20.0 mA. Enter when done to return to system
menu.
The range of adjustment is approxi/- 10% of scale (+/- 1.6 ma).
Since the 4 to 20 mA output is tied to the 0 to 1 volt output, this function can be
used to calibrate the 0 to 1 volt output, if the 4 to 20 mA output is not used. By
using a digital Volt meter on the 0-1 Volt output.
3.3.12 Manual control of filter and solenoids
For troubleshooting purposes, you have manual access to control
calibration filter and solenoid on the Analyze mode. To have manual access to
the calibration filter and solenoid:
Summary of Contents for 6600
Page 4: ...iv Model 6600 Teledyne Analytical Instruments ...
Page 8: ...iv Part I Model 6600 Oil in Water Analyzer Teledyne Analytical Instruments ...
Page 24: ...2 Installation Model 6600 2 10 Part I Teledyne Analytical Instruments ...
Page 64: ...iv Part II Model 6600 Oil in Water Analyzer Teledyne Analytical Instruments ...
Page 86: ...3 Maintenance Model 6600 3 10 Part II Teledyne Analytical Instruments ...
Page 147: ...Part III 1 57 Oil in Water Part III Teledyne Analytical Instruments ...