Oil in Water Analyzer
Operation 3
Part I 3-21
Teledyne Analytical Instruments
3.4.2 Span Cal
The
Span
function on the main menu is used to span calibrate the analyzer.
Span calibration can be performed in either the automatic or manual mode.
Make sure the span fluid is flowing to the instrument.
3.4.2.1
Auto Mode Spanning
Observe all precautions in sections 3.4 and 3.4.2, above. Enter the span
function. The screen that appears allows you to select whether the span calibra-
tion is to be performed automatically or manually. Use the
UP/DOWN
SWITCH
to toggle between
AUTO
and
MAN
span settling. Stop when
AUTO
appears,
blinking, on the display.
Select span
mode: AUTO
Enter
to move to the next screen.
Span Val: 2Ø.ØØ %
<ENT> twice to start
The unit field should be blinking first (%/ppm). Use the UP/DOWN switch
to set the proper unit of the span fluid and enter. Then, use the UP/DOWN
switch to set the concentration. When you have set the concentration of the
span fluid you are using,
Enter
to begin the Span calibration.
####.## ppm
OIL
Slope=#.### Span
The beginning span value is shown in the upper left corner of the display.
As the span reading settles, the screen displays and updates information on
Slope
. Spanning automatically ends when the span output corresponds, within
tolerance, to the value of the span gas concentration. Then the instrument auto-
matically returns to the analyze mode.
3.4.2.2
Manual Mode Spanning
Enter the
Span
function. The screen that appears allows you to select
whether the span calibration is to be performed automatically or manually.
Select span
mode: MANUAL
Summary of Contents for 6600
Page 4: ...iv Model 6600 Teledyne Analytical Instruments ...
Page 8: ...iv Part I Model 6600 Oil in Water Analyzer Teledyne Analytical Instruments ...
Page 24: ...2 Installation Model 6600 2 10 Part I Teledyne Analytical Instruments ...
Page 64: ...iv Part II Model 6600 Oil in Water Analyzer Teledyne Analytical Instruments ...
Page 86: ...3 Maintenance Model 6600 3 10 Part II Teledyne Analytical Instruments ...
Page 147: ...Part III 1 57 Oil in Water Part III Teledyne Analytical Instruments ...