A-2
Appendix
Models 6600
Teledyne Analytical Instruments
Typical Analytical Performance Specifications
:
(
will vary per application
)
Accuracy:
±2% of full scale possible (Oil in Water)
When calibrated on specific oil of interest.
Noise:
Less than ±1%
Drift:
Less than ±1% per auto Zero cycle (source/
detector dependent)
Sample Cell:
(Aluminum/CPVC for Oil in Water) with
Sapphire/Quartz windows standard. (Kynar,
Kalrez optional)
Cell Length:
1/8 to 3.5” (application dependent)
Flow Rate:
100 ccm to 2 lpm (application dependent)
Light Source:
Mercury, Hg
Sensitivity:
.08 to 1.0 absorbance units.
Reproducibility:
+/-1% of scale or better
Filter Wavelength:
210 to 1000 millimicrons. (application depen-
dant - oil in water 254/365mm).
Sample Pressure:
150 psi maximum (oil in water applications)
Sample temperature:
1-120
o
C (34-250
o
F)
(non freezing, non-boiling)