TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 9-1: Test Functions (Continued)
Test Function
Factory Set-Up
Comment
SLOPE
1.0 ± 0.3
The slope can be thought as a gain term which determines the
steepness of the calibration curve.
The 6200A will operate normally even the slope is out of range,
however it is preferred within the range for optimum operation
of the analyzer.
Slope out of range could be due to:
Analog gain pot needs adjustment (see Section 9.1.6).
Poor calibration quality (see Section 7.10).
UV lamp not calibrated (see Section 9.1.6)
OFFSET
<250 mV
This is essentially identical to the stray light except it is
expressed in mV.
High offset could be due to:
Light leak.
Aging UV filter.
Contaminated reaction cell.
Pneumatic leak.
Poor calibration quality.
HVPS
550 - 900 V
This represents the scaled-up HVPS programming voltage to
the HVPS. The design of the HVPS precludes taking a single
reading that indicates the health of the supply. Refer to the
HVPS Troubleshooting Section 9.3.10 for a procedure for
testing the HVPS. This TEST function is used primarily to set
the HVPS voltage value and the reading should be very stable
and constant. A value not in the 400 to 900 volt range indicates
problems with the HVPS supply.
DCPS
2500 ± 200 mV
DCPS is a composite of the +5 and ± 15VDC supplies. It has
been arbitrary set at 2500 ± 200mV. If it is not in this range one
of the voltages in the supply is not working. Check the
procedures for diagnosing the Power Supply Module.
(table continued)
9-7
Summary of Contents for 6200A
Page 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Page 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Page 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Page 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Page 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Page 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Page 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Page 27: ......
Page 35: ......
Page 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Page 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Page 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Page 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Page 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Page 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Page 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Page 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Page 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Page 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Page 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Page 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Page 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...