TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 7-9: Activity Matrix for Calibration Procedure
Equipment/supplies
Acceptance limits
Frequency and method
of measurement
Action if requirements
are not met
Calibration gases
NIST traceable
Assayed against an
NIST-SRM semi-
annually, Sec. 2.0.7,
(Q.A. Handbook)
Working gas standard is
unstable, and/or
measurement method is
out of control; take
corrective action such
as obtaining new
calibration gas.
Dilution gas
Zero air, free of
contaminants
See Section 2.9.2 (Q.A.
Handbook)
Return to supplier or
take appropriate action
with generation system
Multi-point
calibration
Use calibration
procedure in
Subsec. 2.2 (Q.A.
Handbook); also
Federal Register
Perform at least once
every quarter or anytime
a level span check
indicates a discrepancy,
or after maintenance
which may affect the
calibration; Subsec 2.5
(Q.A. Handbook)
Repeat the calibration
7.8.3 Data Recording Device
Either a strip chart recorder, data acquisition system, digital data acquisition system should be
used to record the data from the 6200A RS-232 port or analog outputs. If analog readings are
being used, the response of that system should be checked against a NIST referenced voltage
source or meter. Data recording device should be capable of bi-polar operation so that negative
readings can be recorded.
7.8.4 Dynamic Multipoint Span Calibration
Dynamic calibration involves introducing gas samples of known concentrations to an instrument
in order to adjust the instrument to a predetermined sensitivity and to derive a calibration
relationship. A minimum of three reference points and one zero point uniformly spaced covering
0 to 80 percent of the operating range are recommended to define this relationship.
The analyzer's recorded response is compared with the known concentration to derive the
calibration relationship.
7-12
Summary of Contents for 6200A
Page 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Page 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Page 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Page 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Page 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Page 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Page 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Page 27: ......
Page 35: ......
Page 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Page 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Page 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Page 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Page 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Page 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Page 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Page 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Page 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Page 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Page 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Page 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Page 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...