TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
7.8.9.2 Precision Check
A periodic check is used to assess the data for precision. A one-point precision check must be
carried out at least once every 2 weeks on each analyzer at an SO
2
concentration between 0.08 and
0.10 ppm. The analyzer must be operated in its normal sampling mode, and the precision test gas
must pass through all filters, scrubbers, conditioners, and other components used during normal
ambient sampling. The standards from which precision check test concentrations are obtained must
be traceable to NIST-SRM. Those standards used for calibration or auditing may be used.
7.8.9.3 Precision Check Procedure
1. Connect the analyzer to a precision gas that has an SO
2
concentration between 0.08 and 0.10
ppm. An SO
2
precision gas may be generated by either the SO
2
gas cylinder or a SO
2
permeation tube. If a precision check is made in conjunction with a zero/span check, it must
be made prior to any zero or span adjustments.
2. Allow the analyzer to sample the precision gas until a stable trace is obtained.
3. Record this value. Information from the check procedure is used to assess the precision of the
monitoring data; see 40 CFR 58 for procedures for calculating and reporting precision.
7.8.10 Recommended Standards for Establishing Traceability
To assure data of desired quality, two considerations are essential: (1) the measurement process
must be in statistical control at the time of the measurement and (2) the systematic errors, when
combined with the random variation in the measurement process, must result in a suitably small
uncertainty.
Evidence of good quality data includes documentation of the quality control checks and the
independent audits of the measurement process by recording data on specific forms or on a
quality control chart and by using materials, instruments, and measurement procedures that can
be traced to appropriate standards of reference. To establish traceability, data must be obtained
routinely by repeat measurements of standard reference samples (primary, secondary, and/or
working standards). More specifically, working calibration standards must be traceable to
standards of higher accuracy, such as those listed below in Table 7-15.
7-23
Summary of Contents for 6200A
Page 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Page 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Page 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Page 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Page 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Page 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Page 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Page 27: ......
Page 35: ......
Page 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Page 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Page 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Page 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Page 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Page 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Page 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Page 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Page 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Page 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Page 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Page 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Page 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...