TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
All monitoring instrument systems are subject to some drift and variation in internal parameters
and cannot be expected to maintain accurate calibration over long periods of time. Therefore, it
is necessary to dynamically check the calibration relationship on a predetermined schedule. Zero
and span checks must be used to document that the data remains within control limits. These
checks are also used in data reduction and validation. Table 7-9 summarizes the initial quality
assurance activities for calibrating equipment. Table 7-10 is a matrix for the actual dynamic
calibration procedure.
Calibrations should be carried out at the field monitoring site.
The Analyzer should be in
operation for at least several hours (preferably overnight) before calibration so that it is
fully warmed up and its operation has stabilized.
During the calibration, the 6200A should be
in the CAL mode, and therefore sample the test atmosphere through all components used during
normal ambient sampling and through as much of the ambient air inlet system as is practicable.
If the Instrument will be used on more than one range, it should be calibrated separately
on each applicable range (see Section 7.9).
Calibration documentation should be maintained
with each analyzer and also in a central backup file.
7.8.2 Calibration Gas Sources
Compressed SO
2
in nitrogen
The NIST-SRM's provide references against which all calibration gas mixtures must be
compared (Section 2.0.7, Q.A. Handbook). The procedure requires the comparison of the
concentration of a commercial, working calibration standard to an NIST-SRM. This is described
in Subsection 7.1 of Section 2.0.7, Q.A. Handbook. Subsections 7.1.4 and 7.1.5 describe the
verification and reanalysis of cylinder gases.
SO
2
permeation tubes
The steps required to compare the concentration of a commercial working calibration standard to
an NIST-SRM are described in Subsection 7.2.3 of Section 2.0.7, Q.A. Handbook. See
Subsection 7.2.6 for the re-analysis of permeation tubes.
Dilution air
Zero air (free of contaminants that could cause a detectable response with the Sulfur Dioxide
Analyzer) is commercially available, or can be generated by the user. A clean air system
utilizing ambient air may be more desirable to use for zero and dilution purposes. If compressed
air cylinder is used, the air should have O
2
, N
2
, and CO
2
content similar to that of ambient air
and less than 0.1ppm aromatic hydrocarbons.
7-10
Summary of Contents for 6200A
Page 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Page 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Page 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Page 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Page 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Page 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Page 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Page 27: ......
Page 35: ......
Page 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Page 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Page 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Page 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Page 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Page 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Page 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Page 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Page 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Page 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Page 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Page 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Page 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...