Photometric Analyzer
Operation 3
Part I 3-29
Teledyne Analytical Instruments
tabular data of this type for a large number of gases, which it makes available to
customers on request. See Appendix for ordering information. To enter data:
From the
System
Functions Screen—
1. Use < > to select
ALGORITHM
, and
Enter
.
2. Select and
Enter
SETUP
.
3.
Enter
MANUAL
from the Calibration Mode Select screen.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ
The data entry screen resembles the verify screen, but the gas values can
be modified and the data-point number cannot. Use the
< >
keys to toggle
between the
INPUT
and
OUTPUT
fields. Use the
∆∇
keys to set the value for the
lowest concentration into the first point. Then press
Enter
.
After each point is entered, the data-point number increments to the next
point. Moving from the lowest to the highest concentration, use the
∆∇
keys to
set the proper values at each point.
Dpt INPUT OUTPUT
0 Ø.ØØ Ø.ØØ
Repeat the above procedure for each of the data points you are setting (up
to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The
linearizer will automatically adjust for the number of points entered.
When you are done, Press
ESCAPE
. The message,
Completed. Wait for
calculation,
appears briefly, and then the main
System
screen returns.
To end the session, send:
st<enter>
st<enter>
to the analyzer from the computer.
3.8.2.2 Auto Mode Linearization
To linearize in the Auto Mode, you must have on hand a separate calibra-
tion gas for each of the data points you are going use in your linearization. First,
the analyzer is zeroed and spanned as usual. Then, each special calibration gas,
for each of the intermediate calibration points, is flowed, in turn, through the
sensor. As each gas flows, the differential value for that intermediate calibration
point is entered from the front panel of the analyzer.
Note: The span gas used to span the analyzer must be >90% of the
range being analyzed.
Summary of Contents for 6000A
Page 4: ...iv Model 6000A Teledyne Analytical Instruments...
Page 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Page 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...