![Teledyne 6000A Operating Instructions Manual Download Page 39](http://html1.mh-extra.com/html/teledyne/6000a/6000a_operating-instructions-manual_1080454039.webp)
Photometric Analyzer
Operation 3
Part I 3-15
Teledyne Analytical Instruments
OUTPUT: 4 MA and OUTPUT: 20 MA can be toggled by moving on
that field and pressing the Up or Down key. 4 mA output should be calibrat-
ed first and 20 mA output afterwards.
4.
Select OUTPUT: 4 MA and press the Enter key
Use UP/DOWN arrow to
Adjust 4 ma: 250
The number on the second row is the setpoint of the 4 mA output. It is
analogous to a potentiometer wiper. The number can be set anywhere from 0
to 500. The default is 250, in the middle. At the default setting, the output
should be very close to 4 mA. If not, slowly adjust the number using the Up or
the Down keys until DMM reads 4.00 mA. Press the Enter key when done.
5.
Now select OUTPUT: 20 MA and press the Enter key. A
screen similar to the one above will appear and the DMM should read close to
20 mA. If not, slowly adjust the number using the Up or Down key until DMM
reads 20.0 mA. Press the Enter key when done.
The range of adjustment is approxi/- 10% of scale (+/- 1.6 ma).
Since the 4 to 20 mA output is tied to the 0 to 1 volt output, this function can be
used to calibrate the 0 to 1 volt output, if the 4 to 20 mA output is not used. By
using a digital Volt meter on the 0-1 Volt output.
3.3.12 Model
This selection in the System menu flashes for a few seconds the model
number and the software version installed in this instrument.
3.4
The
Zero and Span Functions
The Model 6000A can have as many as three analysis ranges plus a
special calibration range (Cal Range). Calibrating any one of the ranges will
automatically calibrate the other ranges.
CAUTION: Always allow 4-5 hours warm-up time before calibrat-
ing, if your analyzer has been disconnected from its
power source. This does not apply if the analyzer
was plugged in but was in STANDBY.
The analyzer is calibrated using zero, and span gases.
Note: Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit pressure to 40 psig or less when turning it
back on.
Summary of Contents for 6000A
Page 4: ...iv Model 6000A Teledyne Analytical Instruments...
Page 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Page 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...