3 Operation
Model 6000A
3-14 Part I
Teledyne Analytical Instruments
AUTOCAL FILSOL HOLD
CAL-HOLD-TIMER MORE
3.
Select with the Right or Left keys CAL-HOLD-TIMER, and
press the Enter key to access this function menu:
Calbrt hold:
3 min
Sample hold: 1 min
The calibration hold time is set on the first row, while the sample hold
time is set on the second row. To select one or the other, use the Right or
Left keys. To modify the time of either timer, use the Up or Down keys.
The time is in the minutes.
3.3.11 Analog 4 to 20 mA Output Calibration
This function will let the operator calibrate the 4 to 20 mA analog output to
match the display reading. A DMM configure as a DC ammeter is needed. The
DMM should be connected across the output terminals of the 4 to 20 mA output
to monitor the output current. To enter the 4 to 20 mA output adjust function,
you must:
1.
Press the System key to start the System function:
DIG_FILT SELF-TEST
PWD LOGOUT MORE
2.
Using the Right or Left arrow keys, select MORE and press
Enter. The second System screen appears:
AUTOCAL FILSOL TRACK
CAL-HOLD-TIMER MORE
or
AUTOCAL FILSOL HOLD
CAL-HOLD-TIMER MORE
3.
Using the Right or the Left arrow keys, select MORE and
press Enter. The third System screen appears:
ALGORITHM APPLICATION
MODEL OUTPUT: 4 MA
or
ALGORITHM APPLICATION
MODEL OUTPUT: 20 MA
Summary of Contents for 6000A
Page 4: ...iv Model 6000A Teledyne Analytical Instruments...
Page 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Page 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Page 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...