background image

Photometric Analyzer

Operation  3

Part I    3-19

Teledyne Analytical Instruments

Use the < > arrow keys to toggle between the span concentration value

and the units field (%/ppm). Use the 

∆∆∆∆∆∇

  arrow keys change the value and/or the

units, as necessary. When you have set the concentration of the span gas you are
using, press 

Enter 

to begin the Span calibration.

  ####.##%

   SO2

Slope=#.###     Span

The beginning span value is shown in the upper left corner of the display.

As the span reading settles, the screen displays and updates information on

Slope

. Spanning automatically ends when the span output corresponds, within

tolerance, to the value of the span gas concentration. Then the instrument auto-
matically returns to the analyze mode.

3.4.2.2

Manual Mode Spanning

Press 

Span

 to start the 

Span

 function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Select span

mode:  MANUAL

Use the 

∆∇

 keys to toggle between 

AUTO

 and 

MAN

 span settling. Stop

when 

MAN

 appears, blinking, on the display. Press 

Enter

 to move to the next

screen.

Span Val:  2Ø.ØØ %

<ENT> To begin span

Use the < > arrow keys to toggle between the span concentration value

and the units field (%/ppm). Use the 

∆∆∆∆∆∇

  arrow keys change the value and/or the

units, as necessary. When you have set the concentration of the span gas you are
using, press 

Enter 

to begin the Span calibration.

Press 

Enter

 to enter the span value into the system and begin the span

calibration.

Once the span has begun, the microprocessor samples the output at a

predetermined rate. It calculates the difference between successive samplings
and displays this difference as Slope on the screen. It takes several seconds for
the first Slope value to display. Slope indicates rate of change of the Span
reading. It is a sensitive indicator of stability.

####.##%        SO2

Slope=#.###     Span

Summary of Contents for 6000A

Page 1: ...YSTEM PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF...

Page 2: ...t of any parts in so far as such alterations do not adversely affect our warranty ImportantNotice This instrument provides measurement readings to its user and serves as a tool by which valuable data...

Page 3: ...iii Photometric Analyzer Teledyne Analytical Instruments Table of Contents Specific Model Information iv Preface v Part I Control Unit Part I 1 1 Part II Analysis Unit Part II 1 1 Appendix A 1...

Page 4: ...iv Model 6000A Teledyne Analytical Instruments...

Page 5: ...Part I i Part I Control Unit Teledyne Analytical Instruments OPERATING INSTRUCTIONS Model 6000A Photometric Analyzer Part I Control Unit Panel Mount...

Page 6: ...System 2 10 3 Operation 3 1 Introduction 3 1 3 2 Using the Data Entry and Function Buttons 3 1 3 3 The System Function 3 4 3 3 1 Setting up an Auto Cal 3 4 3 3 2 Password Protection 3 6 4 3 2 1 Enter...

Page 7: ...ms Function 3 20 3 6 The Range Select Function 3 22 3 6 1 Manual Select Define Range Screen 3 23 3 6 2 Auto Screen 3 23 3 6 3 Precautions 3 24 3 7 The Analyze Function 3 25 3 8 Programming 3 26 3 8 1...

Page 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...

Page 9: ...a nonhazardous environment only The Analysis Unit or Remote Section can be designed for a variety ofhazardousenvironments 1 2 Typical Applications A few typical applications of the Model 6000A are Oi...

Page 10: ...ally select the proper preset range for a given measurement Manual override allows the user to lock onto a specific range of interest Twoadjustableconcentrationalarmsandasystemfailurealarm Extensive s...

Page 11: ...e used to change the specific function performed by the analyzer Analyze Perform analysis for concentration content of a sample System Performsystem relatedtasks describedindetailinchapter 3 Operation...

Page 12: ...screen is an easy to use interface between operator and analyzer It displays values options and messages for immediate feedback to the operator Power Button The button switches the instrument power b...

Page 13: ...Part I 1 5 Photometric Analyzer Part I Control Unit Teledyne Analytical Instruments Figure 1 2 Model 6000A Rear Panel...

Page 14: ...mote Span Zero Digital inputs allow external control of analyzercalibration Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample...

Page 15: ...ered in Part II of this manual 2 1 Unpacking the Control Unit The analyzer is shipped with all the materials you need to install and prepare the system for operation Carefully unpack the Control Unit...

Page 16: ...ntrols are mounted on the inner control panel door which is hinged on the left edge and doubles as a door to provide access to the internal components of the instrument The door will swing open when t...

Page 17: ...wer communications and both digital and analog signal outputs are described in the following paragraphs Wire size and maximum length data appear in the Drawings at the back of this manual Figure 2 3 I...

Page 18: ...eceptacle is not used Fuses are not installed at the factory Be sure to install the proper fuse as part of installation See Fuse Replace ment in chapter 5 maintenance Analog Outputs There are four ter...

Page 19: ...rrently selected analysis range For example if the analyzer is set on a range that was defined as 0 10 then the output would be as shown in Table 2 1 Table 2 1 Analog Concentration Output Examples Ana...

Page 20: ...acts Each provides a set of Form C contacts for each type of alarm Each has both normally open and normally closed contact connections The contact connections are indicated by diagrams on the rear pan...

Page 21: ...erminal block They accept 0 V OFF or 24 V dc ON for remote control of calibration See Remote Calibration Protocol below Zero Floating input 5 to 24 V input across the and terminals puts the analyzer i...

Page 22: ...remove the zero command 3 When CRC opens again send a span command until the CRC closes The CRC will quickly close 4 When the CRC closes remove the span command When CRC opens again zero and span are...

Page 23: ...range 0 100 etc Which alarm if any are disabled AL x DISABLED Which alarms if any are tripped AL x ON Each status output is followed by a carriage return and line feed Three input functions using RS 2...

Page 24: ...luded in the system 2 4 Testing the System After The Control Unit and the Analysis Unit are both installed and interconnected and the system gas and electrical connections are complete the system is r...

Page 25: ...tion cycle if desired RoutineOperation Calibratetheinstrument Choose autoranging or select a fixed range of analysis Set alarm setpoints and modes of alarm operation latching fail safe etc Program Rep...

Page 26: ...iliary tasks Checking model and software version Adjust electronic filter of the signal Displaymoresubfunctions Two of these are for programming reprogramming the analyzer Define gas applications and...

Page 27: ...CAL Enter Password Define Application Range Select range Display gas use andrange Enter Secure System setup not allowed Span Zero status and setup Span Zero timing and on off Enter Span Zero Solenoid...

Page 28: ...timer schedule for Zero and Span cycling FILSOL Select Span Zero flag filter or Span Zero solenoid valveforcalibrationmethod TRACK Setthesystemreadingtobeheldorfollowedbythe concentration gasorfilter...

Page 29: ...blink ZERO or SPAN then press Enter again You won t be able to set OFF to ON if a zero interval is entered A Span Every or Zero Every screen appears Zero schedule OFF Day d Hour h Use arrows to set a...

Page 30: ...the password in a separate safe location 3 3 2 1 EnteringthePassword Toinstallanewpasswordorchangeapreviouslyinstalledpassword you mustkeyinandENTERtheoldpasswordfirst Ifthedefaultpasswordisin effect...

Page 31: ...edefaultTAIortheprevious lyassignedpassword orpressEscapetokeeptheexistingpasswordand moveon IfyouchoseEntertochangethepassword thepasswordassignment screenappears Selectnewpassword T A I Enter the pa...

Page 32: ...videsaconvenientmeansofleavingtheanalyzer inapasswordprotectedmodewithouthavingtoshuttheinstrumentoff By enteringLOGOUT youeffectivelylogofftheinstrumentleavingthesystem protectedagainstuseuntilthepas...

Page 33: ...NNING DIAGNOSTIC Testing Preamp Cell Whenthetestingiscomplete theresultsaredisplayed Power OK Analog OK Cell 2 Preamp 3 ThemoduleisfunctioningproperlyifitisfollowedbyOK Anumber indicatesaprobleminaspe...

Page 34: ...to return to the previous screen Select and Enter SET UP toCalibrationModescreen set upwillnotworkwithoutaPCbeingconnected totheanalyzer Selectalgorithm mode AUTO Therearetwowaystolinearize AUTOandMA...

Page 35: ...deoffinnoise isofvalue theoperatorcoulddecreasethenumberofthedigitalfilter In applicationswherethesignalisnoisy theoperatorcouldswitchtoahigher number theresponsetimeissloweddownthough 90 responsetime...

Page 36: ...rationandtheotherforSpanlocatedintheDetector housing TomovebetweentheZeroandtheSpanflags usetheRightorLeft arrowkeys FILmeansthatafilterwilldothisparticularcalibration SOL meansthatthesignaltoactivate...

Page 37: ...ingcalibration The analog outputs are both 0 to 1 volt outputs and both 4 to 20 mA outputs 3 3 10 Calibration Hold Timer Setup ThisCalibrationTimerletstheoperatoradjustthetimetheeinstrument purgesthec...

Page 38: ...put Calibration Thisfunctionwilllettheoperatorcalibratethe4to20mAanalogoutputto matchthedisplayreading ADMMconfigureasaDCammeterisneeded The DMMshouldbeconnectedacrosstheoutputterminalsofthe4to20mAout...

Page 39: ...gtheUporDownkeyuntilDMM reads 20 0 mA Press the Enter key when done Therangeofadjustmentisapproximately 10 ofscale 1 6ma Sincethe4to20mAoutputistiedtothe0to1voltoutput thisfunctioncanbe used to calibr...

Page 40: ...wsyoutoselectwhetherthezero calibrationistobeperformedautomaticallyormanually Usethe arrow keystotogglebetweenAUTOandMANzerosettling StopwhenAUTOappears blinking onthedisplay Selectzero mode AUTO Pres...

Page 41: ...tof threezeroingscreensappears Thenumberintheupperlefthandcorneristhe first stagezerooffset Themicroprocessorsamplestheoutputatapredeter minedrate SO2 Zero adj 2048 C Zero TheanalyzergoesthroughC Zero...

Page 42: ...the flow is no more than 200 SCCM d Checktemperaturecontrollerboard e Check gas temperature f Check the Sample Cell for dirty windows Ifnoneoftheabove proceedtoperformanopticalbalanceasdescribedin se...

Page 43: ...atappearsallows youtoselectwhetherthespancalibrationistobeperformedautomaticallyor manually Selectspan mode MANUAL Usethe keystotogglebetweenAUTOandMANspansettling Stop whenMANappears blinking onthedi...

Page 44: ...failsafe or non failsafe or they can be defeated altogether Thesetpointsforthealarmsarealsoestablishedusingthisfunction Decidehowyouralarmsshouldbeconfigured Thechoicewilldepend uponyourprocess Consid...

Page 45: ...oproceed entertheAlarmfunction PresstheAlarmbuttononthefrontpaneltoentertheAlarmfunction Use the keys to choose between or ppm units Then press Enter to movetothenextscreen AL1 1 ppm HI Dft N Fs N Ltc...

Page 46: ...the System function if linearization is to apply through out the range Furthermore if the limits are too small a part approx 10 or less of the originally linear ized range the linearization will be c...

Page 47: ...ngeIDoutputs willbe noticed Theautorangingfeaturecanbeoverriddensothatanalogoutputstays ona fixedrangeregardlessofthecontaminantconcentrationdetected Iftheconcen trationexceedstheupperlimitoftherange...

Page 48: ...r limits but upper limits that differ by approximately an order of magnitude Ranges whose upper limits coincide with the lower limits of the next higher range Ranges where there is a gap between the u...

Page 49: ...the EscapebuttoninmanycasesalsoswitchestheanalyzerbacktotheAnalyze function Alternatively youcanpresstheAnalyzebuttonatanytimetoreturn toanalyzingyoursample TheAnalyzefunctionscreenshowstheimpuritycon...

Page 50: ...ion for details Send the rp command to the analyzer 3 Press the System button to start the System function DIG_FILT SELF TEST PWD LOGOUT MORE Use the arrow keys to blink MORE then press Enter AUTOCAL...

Page 51: ...allstobelow85 offullscaleof thenextlowerrange theinstrumentswitchestothelowerrange Acorrespond ingshiftintheDCconcentrationoutput andintherangeIDoutputs willbe noticed Theautorangingfeaturecanbeoverri...

Page 52: ...s established by the operator for that particular range the output saturates at 1 V dc or 20 mA However the digital readout and the RS 232 output continue to read regardless of the analog output range...

Page 53: ...eysto setthepropervaluesateachpoint Dpt INPUT OUTPUT 0 Repeattheaboveprocedureforeachofthedatapointsyouaresetting up toninepoints 0 8 Setthepointsinunitincrements Donotskipnumbers The linearizerwillau...

Page 54: ...O from the Calibration Mode Select screen TheAutoLinearizeModedataentryscreenappears 19 5 ppm SO2 Input 20 00 5 Use the keys to set the proper value of calibration gas and Enter Repeat this step for e...

Page 55: ...and fuses andrecalibration WARNING SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL 4 1 Fuse Replacement 1 Place small screwdriver in notch and pry cover off as shown in Figure 4 1 Figure 4 1 Removing F...

Page 56: ...system mode 2 Use the arrow keys to move to More and press Enter 3 Use the arrow keys to move to Self Test and press Enter The following failure codes apply Table 4 1 Self Test Failure Codes Power 0...

Page 57: ...ontrol Units contain the following major components Power Supply Motherboard with Microprocessor RS 232 chip and PreamplifierPCB Front Panel Display Board and Displays 5 digit LED meter 2 line 20 char...

Page 58: ...connect the instru ment from the power source Close and latch the front panel access door Clean outside surfaces with a soft cloth dampened slightly with plain clean water Do not use any harsh solvent...

Page 59: ...uments OPERATING INSTRUCTIONS Model 6000A Photometric Analyzer Part II Analysis Unit NEC Type Part Number D 65478 6000A GP Rack Panel Integral or Remote 6000B GP Bulkhead Z Purged in Div II areas Inte...

Page 60: ...king the Analyzer 1 1 2 2 Installing Connecting the Analyzer 1 1 2 2 1 User Connections 3 1 2 2 2 Electrical Power Connections 1 2 2 2 3 Compressed Air Supply 2 2 2 2 4 Pipe Connections 2 2 2 2 5 Sign...

Page 61: ...of the I to E Converter Output 3 4 3 5 5 Balancing the Optics for Equal Light Transmission with Zero Fluid in the Sample Cell 3 5 3 5 6 Setup of the Logarithmic Amplifier 3 6 3 5 7 Inverting Amplifier...

Page 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...

Page 63: ...s the course of optical energy in the analyzer The optical energy is emitted from a source lamp in the source module passed through the sample cell and received by the sensor which converts the optica...

Page 64: ...optical filters mounted 180 degrees apart that alternately and continuously rotate into and out of the light beam Sample flows continuously through the sample cell and absorbs optical energy at vario...

Page 65: ...Part II 1 3 Photometric Analyzer Operational Theory 1 Teledyne Analytical Instruments 6000A Unit Only D 69023...

Page 66: ...1 4 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments 6000B Remote Control Unit 6000B Integral Control Unit...

Page 67: ...nterest the output of the I to E converting amplifier is fed to the input of a logarithmic amplifier which produces a signal that represents the logarithm of the output signal of the second amplifier...

Page 68: ...it is a digital circuit which employs a DAC Digital to Analog Converter that can go out of range When the threshold cannot be found oscillation persists this means that measuring and reference peak si...

Page 69: ...Description Thephotometricanalyzerisgenerallyconstructedforgeneral purpose Model 6000A use and is mounted on a BACKPLATE an open rack or in a closedcubicle 1 6 Photometer Thethreephotometermodulesare...

Page 70: ...tains a constant anode current in the D2 lamp and controls the voltage across the lamp s cathode filament When power is turned on relay K1 is activated and applies 10 VDC across the filaments After io...

Page 71: ...u must align a reference mark on the wheel with a reference mark on the shaft When the switch activating disc is removed align with the marks on the switch plate and motor mount when you put it back T...

Page 72: ...ctor and Preamplifier PCB 1 7 Sample Systems Below are sample systems that deliver gases to the 6000 6020 sample cell of the Analysis Unit Depending on the mode of operation either sample or calibrati...

Page 73: ...V 51 F 383 MODEL 6000 ANALYZER Hi pressure reduction 0 50 psig for faster response Sample in Inst Air in 90 120 psig V 469 F 1242 Sample Return Eductor pump By Pass Flow 2 20 SCFH Zero in Span in G 3...

Page 74: ...CELL 100C BACK PRESSURE NOT REQURIED IF RETURN IS STABLE AT 1 0 KG CM2A Span in Differential pressure regulator 3 PSID Fast loop and By Pass Filter 5 microns Zero in By Pass Flowmeter 2 20 GPH Sample...

Page 75: ...dyne Analytical Instruments Model 6000B Photometric Analyzer with D2 Lamp Model 6000B Photometric Analyzer with D2 Lamp Analysis Bench shown with Integral General purpose bulkhead Control Unit Control...

Page 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...

Page 77: ...erature Control the system must be installed in an area where the ambient temperature is not permitted to drop below 32 F nor rise above 110 F Regardlessofconfiguration thesystemmustbeinstalledonaleve...

Page 78: ...ut pipe connec tions On special systems consult the text in the manual that describes your particularsamplesystemindetail 2 2 5 Signal and Alarm Output Connections Signal and alarm output connections...

Page 79: ...ments of your process Power up the system and test it by performing the following operations 1 Repeat the Self Diagnostic Test 2 4 Calibration 2 4 1 Calibration Fluids Zeroandspanfluidsmustbemadebythe...

Page 80: ...ection I 2 Injectspanfluidandsettheconcentrationofthespanfluidwith the span procedure referred in section 3 4 section I Method Two 1 Determine the span setting using Method One 2 Activate the span fil...

Page 81: ...the lamp circuit or its starter Youcandetectalampoffconditionwiththesignalfailurealarmcircuit butyoumustconnecttherelaycontactsfromthealarmtoyourindicatingdevice In addition you will experience an ala...

Page 82: ...ures empirically because the duration of time that the system runs without attention is related directly to the sample scondition 3 4 Suggested Preventive Maintenance Schedule DAILY 1 Visuallyinspectt...

Page 83: ...to AC to observe the demodulator switch signals DVM DigitalVoltmeter PC Board Extender Use the PC board extender whenever you need to adjust trimpot Because all PC board connectors are keyed to avoid...

Page 84: ...analyzer keeps operating within its dynamic range ConnecttheoscilloscopetoTP3 Theoscilloscopedisplaysthemeasur ing and reference pulses in an alternating pattern The display is created by the light pa...

Page 85: ...l R2onthedetectorPCboard Whenthiscannotbedonebecausebothpeaksare tooshortortoolong searchforscreensmountedinthelight path usuallylocated inaholderonthelightpipewhichinterconnectsthedetectorandsamplemo...

Page 86: ...tinuously taking the logarithm of the outputsignalofthesecondamplifier Youcanobservetheoutputbyconnecting the scope probe to TP4 The correct wave shape has a rounded negative going pulse that is the s...

Page 87: ...t and a rotating chopper disc activate themagneticmercuryreedswitch Analuminummotormountingblockhouses a bar magnet This bar magnet is parallel with the mercury chopper switch The chopper disc is a gr...

Page 88: ...The labels stand for Neutral Ground and Hot This is the AC power strip terminal It feeds AC power to other components of the Model 6000B System such as the D2 lamp power supply heater and temperature...

Page 89: ...the photodetector DC power for the photodetector preamplifier and control signals for the sole noids and filters The labels stand for 230 VDC This is the negative high voltage fed to the photodetector...

Page 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...

Page 91: ...Display 2 line by 20 alphanumeric VFD accompanied by 5 digit LED display SignalOutput Two 0 1V DC concentration and range ID Two 4 20mADC isolated concentration and range ID RS232 Alarm Two fully pro...

Page 92: ...epen dent Diurnal Less than 1 per 20o F source detector dependent SampleCell Stainless steel with Quartz window standard Othermaterialsavailable Cell Length 01 to 40 inches Flow Rate 50 to 1500 cc min...

Page 93: ...roller SampleCell 1 L 179 Source Lamp 5 F 57 Fuse 5A Slo Blo 5 F 14 Fuse 10A Slo Blo 1 P 43 Phototube 2 C87 Sample Cell Window Quartz 6 O81 Viton O Ring 1 A 16776 Accessory Kit 2 F1295 Fuse 4A slo blo...

Page 94: ...67998 Amplifier PCB Schematic A 38750 Phototube Power Supply Schematic B 36470 DetectorModule PhototubeSchematic B 37533 Detector ConverterPCB PhototubeSchematic B 55819 D2 Power Supply Module Schemat...

Reviews: