4 Operation
Model 3010
PA
4-14: Part I
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in Section
4.3.3 to enter your password. Once you have clearance to proceed, enter the
Alarm
function.
Press the
Alarm
button on the front panel to enter the
Alarm
function.
Make sure that
AL–1
is blinking.
AL1 AL2
Choose Alarm
Set up alarm 1 by moving the blinking over to
AL–1
using the < >
arrow keys. Then press
Enter
to move to the next screen.
AL1 10 % HI
DftN FsN LtchN
Five parameters can be changed on this screen:
•
Value of the alarm setpoint, AL–1 #### (% oxygen)
•
Out-of-range direction, HI or LO
•
Defeated? Dft–Y/N (Yes/No)
•
Failsafe? Fs–Y/N (Yes/No)
•
Latching? Ltch–Y/N (Yes/No).
•
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the
∆∇
arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, setpoint units are
parts-per-million.)
•
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the
∆∇
arrow
keys to change the parameter.
•
Once the parameters for alarm 1 have been set, press
Alarms
again, and repeat this procedure for alarm 2 (AL–2).
•
To reset a latched alarm, go to
Dft–
and then press either
∆
two
times or
∇
two times. (Toggle it to
Y
and then back to
N.
)
–OR –
Summary of Contents for 3010PA
Page 17: ...1 6 Part I 1 Introduction Model 3010PA ...
Page 21: ...2 4 Part I 2 Operational Theory Model 3010PA ...
Page 57: ...iv Part II Model 3010P Oxygen Analyzer ...
Page 69: ...2 8 Part II 2 Operational Theory Model 3010P ...
Page 77: ...3 Installation Model 3010P 3 8 Part II ...