Oxygen Analyzer
Part I: Control Unit
Part I: 4-1
Operation
4.1
Introduction
Once the analyzer has been installed, configure it for your process. To
do this you can:
•
Set system parameters—
•
Specify a password, if desired, requiring operator to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
Calibrate the instrument.
•
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
•
Set alarm setpoints, and modes (latching, failsafe, etc).
Before configuration these default values are in effect:
PARAMETER
DEFAULT
LO Range
1 %
MED Range
5 %
HI Range
10 %
Auto Ranging
ON
Alarm Relays
10 %
(Defeated, HI, Not failsafe, Not latching)
Span
20.9 %
(Auto, every 0 days at 0 hours)
Zero
(Auto, every 0 days at 0 hours).
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press
Enter
for access to all functions of the analyzer.