![Teledyne 3010PA Operating Instructions Manual Download Page 88](http://html1.mh-extra.com/html/teledyne/3010pa/3010pa_operating-instructions-manual_1080412088.webp)
Oxygen Analyzers
Part II: Analysis Units
Part II: 5-7
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few milli-
meters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-3.
Figure 5-3: Removing Fuse Cap and fuse from Holder
2. Replace fuse by reversing process in step 1.
5.4
System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the
< >
arrow keys to move to
More
, and press
Enter
.
3. Use the
< >
arrow keys to move to
Self-Test
, and press
Enter
.
4. Observe the error-code readings on the VFD Display screen,
and check Table 5-1, below, to interpret the codes.