Appendix
3000TA-XL-EU
Teledyne Analytical Instruments
88
damage may occur. For cell pressures above atmospheric, caution must
be exercised to avoid over pressuring the cell holder.
FLOW CONTROL VALVE:
All 3000XL series analyzers have a flow control device. This
device is a needle valve, and serves two functions in the sample path:
1.
It limits the flow rate of the sample through the analyzer. The
needle valve is chosen to operate over a specific range of
pressures and provide a useable flow rate over that range.
2.
The needle valve provides a pressure drop. This device is
selected to provide the only significant pressure drop in the
sample path.
BYPASS:
To improve the system response, a bypass can be added to increase
the sample flow rate to the analyzer by a factor of ten. A by-pass
provides a sample flow path around the analyzer of 2 - 18 SCFH
typically.
CONVERSIONS:
1 PSI
=
2.04 INCHES OF MERCURY (in. Hg.)
1 SCFH =
0.476 SLPM
Summary of Contents for 3000TA-XL-EU
Page 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Page 16: ......
Page 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Page 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Page 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Page 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...
Page 108: ...Appendix 3000TA XL EU Teledyne Analytical Instruments 92...