Operation
3000TA-XL-EU
Teledyne Analytical Instruments
48
4.3.6 Version Screen
Move the ◄► arrow key to More and press
Enter
. With
Version
blinking, press
Enter
. The screen displays the manufacturer, model, and
software version information.
4.4 Calibration of the Analyzer
The analyzer must be calibrated prior to its use. The following
sections describe how to calibrate the analyzer. If equipped with remote
calibration valves, the analyzer can be calibrated from a remote terminal
using the RS-232 port. See Section 3.3.2.3. In addition, web enabled
instruments can set calibration parameters including autocal programs as
well as initiate calibration events remotely from a computer if equipped
with remote calibration valves. See Section 5.2 for more details.
For most applications where the desired range of measurement is 0
to 1 ppm, we recommend the analyzer be calibrated using a span gas as
detailed below:
Span Gas Calibration
Before the cell is ready for calibration, it must be purged with
sample gas to low oxygen levels preferably below 0.2 ppm. However, if
the oxygen content of the sample gas is higher than 0.2 ppm, a zero gas
such as nitrogen with oxygen levels below 0.2 ppm may be required.
The recommended span gas concentration is between 7.0 to 9.0
ppm oxygen in nitrogen, and will require calibration be performed in the
0-10 ppm analyzer range.
Air Calibration
If desired, a newly installed cell may be calibrated in air to 209,000
ppm. However, this practice is generally not recommended as the cell
may take 24 hours or longer to recover to less than 0.2 ppm after being
exposed to air for about 3 to 5 minutes.
Note: The recovery time is significantly less, typically 30 minutes
or less, when the Insta-Trace sensor is used. Refer to
Section 3.4 for information on retrofitting your system to
use an Insta-Trace sensor.
Summary of Contents for 3000TA-XL-EU
Page 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Page 16: ......
Page 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Page 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Page 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Page 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...
Page 108: ...Appendix 3000TA XL EU Teledyne Analytical Instruments 92...