Operation
3000TA-XL-EU
Teledyne Analytical Instruments
42
The factory default is three minutes, but the delay time is
programmable. To adjust to delay time use the
Left
or
Right
arrow keys.
When the time displayed on the second line blinks, it can be adjusted by
Pressing the
Up
or
Down
keys to increase or decrease its value. The
minimum delay is 1 minute, the maximum is 30.
This preference is stored in non-volatile memory so that it is
recovered if power is removed from the instrument.
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see Chapter 3,
Installation
), the analyzer can cycle itself through a sequence of steps
that automatically calibrates the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000TA-XL-EU is acurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
CAUTION:
TAI DOES NOT RECOMMEND FREQUENT ZERO
ADJUSTMENTS OF THE CELL. A NEWLY
INSTALLED CELL MAY TAKE 7-10 DAYS OF
OPERATION TO REACH A STEADY ZERO
(TYPICALLY LESS THAN 0.2 PPM). IF REQUIRED,
THE INSTRUMENT MAY BE ZEROED AFTER THIS
INITIAL STABILIZING PERIOD AND MAY BE
CHECKED AGAIN AFTER A ADDITIONAL 7-10 DAYS.
FREQUENCY OF ZERO ADJUSTMENT IS AT THE
DISCRETION OF THE USER (ONCE A MONTH IS
SUGGESTED).
Choose
System
from the Function buttons. The LCD will display
five sub functions.
TRAK/HLD Auto—Cal
PSWD Logout More
Use ◄► arrows to blink
Auto—Cal
, and press
Enter
. A new screen
for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Summary of Contents for 3000TA-XL-EU
Page 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Page 16: ......
Page 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Page 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Page 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Page 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...
Page 108: ...Appendix 3000TA XL EU Teledyne Analytical Instruments 92...