Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
55
request cells that have been specifically tested in hydrogen streams.
Consult Teledyne for further details.
4.6 The Alarms Function
The Model 3000TA-XL-EU is equipped with 2 fully adjustable
concentration alarms and a system failure alarm. Each alarm has a relay
with a set of form ―C" contacts rated for 3 amperes resistive load at 250
VAC. See Figure 3-5 in Chapter 3,
Installation
and/or the
Interconnection Diagram included at the back of this manual for relay
terminal connections.
The system failure alarm has a fixed configuration as described in
Chapter 3
Installation
.
The concentration alarms can be configured from the front panel as
either
high
or
low
alarms by the operator. The alarm modes can be set as
latching
or
non-latching
, and either
failsafe
or
non-failsafe
, or, they can
be
defeated
altogether. The setpoints for the alarms are also established
using this function.
Decide how your alarms should be configured. The choice will
depend upon your process. Consider the following points:
1.
Which if any of the alarms are to be high alarms and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the
oxygen concentration rises above the setpoint.
Setting an alarm as LOW triggers the alarm when the
oxygen concentration falls below the setpoint.
2.
Decide whether you want the alarms to be set as:
Both high (high and high-high) alarms, or
One high and one low alarm, or
Both low (low and low-low) alarms.
3.
Are either or both of the alarms to be configured as failsafe?
You can set either or both of the concentration alarms to
operate in failsafe or non-failsafe mode.
In failsafe mode, the alarm relay de-energizes in an
alarm condition.
Summary of Contents for 3000TA-XL-EU
Page 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Page 16: ......
Page 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Page 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Page 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Page 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...
Page 108: ...Appendix 3000TA XL EU Teledyne Analytical Instruments 92...