Versa User Manual
www.teledynetekmar.com
how to align the sample/elevator mixer assembly
1. Make sure the mixer wheel is flush with the chassis surface. Use a straight edge to verify.
2. Remove the right-bottom side panel of the unit and set aside.
3. Remove all the vials and the carousel from the top of the Versa. Make sure the Communication Cable is connected.
4. Make sure the unit’s power switch is in the
on
position.
5. Go to diagnostics in Versa TekLink™2G and move the carousel to any position except home.
6. Raise the Sample/Elevator Mixer Assembly to mix.
7. Locate the single beam optical sensor that is mounted on a bracket to the right of the elevator belt.
8. Locate the two Phillips head screws that secure the bracket and adjust the optical sensor in the direction required.
9. Tighten 2 screws and home the Sample/Elevator Mixer Assembly.
10. Visually verify that the mixer wheel is flush with chassis.
11. If all the positions look good, you are done.
12. If it is not flush, slightly adjust the sensor in the correct direction. If you get drastically off, restart from the
beginning.
13. Install the Sample/Elevator Mixer Assembly, carousel and vials back to their original position when all adjustments
are completed.
If any problem is not resolved, contact Teledyne Tekmar Customer Support Center.
retUrning the Versa
do not return the Versa
unless a Teledyne Tekmar representative authorizes you to do so. A service representative
may be able to help you solve the problem over the telephone. If the instrument must be returned, the
representative can tell you how to prevent damage during shipment.
the representative must give you a return
authorization number.
Write this number on the return label.
Unit and Parts disPosaL
do not disCard!
Please contact Teledyne Tekmar or your local distributor for instructions on returning the system for proper
disassembly and disposal.
teLedyne teKmar CUstomer sUPPort:
• U.s. (800) 874-2004
• outside the U.s. (Country Code) + 1 (513) 229-7000
This should be performed by a trained Service Engineer due
to electrical shock hazard.
Maintenance & Troubleshooting • Page 7-12
Summary of Contents for 15-0800-074
Page 1: ...Versa User Manual www teledynetekmar com Versa Manual Part 15 0800 074 ...
Page 8: ...Versa User Manual www teledynetekmar com Chapter 1 Introduction ...
Page 21: ...Versa User Manual www teledynetekmar com Chapter 2 Installation Setup ...
Page 36: ...Versa User Manual www teledynetekmar com Chapter 3 Basic Operation ...
Page 44: ...Versa User Manual www teledynetekmar com Chapter 4 Versa TekLink 2G ...
Page 61: ...Versa User Manual www teledynetekmar com Chapter 5 21 CFR Part 11 Compliance ...
Page 77: ...Versa User Manual www teledynetekmar com Chapter 7 Maintenance Troubleshooting ...
Page 90: ...Versa User Manual www teledynetekmar com Chapter 8 Diagrams ...
Page 91: ...Versa User Manual www teledynetekmar com Detailed Plumbing Diagrams Page 8 2 ...
Page 100: ...Versa User Manual www teledynetekmar com Chapter 9 Index ...