Versa User Manual
www.teledynetekmar.com
The FET gas phase concentration is expressed as follows:
[fet]=
total analyte mass injected into the vial
total gas volume of the vial + net gas phase expansion volume
The analyte mass injected into the vial is determined from the known concentration of standard solutions or the
density for pure compounds. The gas volume of the vial is dependent on the vial size.
NOTE:
Teledyne Tekmar vial volume is 22.3mL (+/- 0.09% RSD).
When preparing a vial for analysis by FET consider the following:
• Final volume of the solvent in the vial should be < 20μl.
• Temperature of the sample platen should be approximately 10-20 degrees higher than the boiling point of the
solvent used to dilute the standard.
• Spiking a vial for FET analysis can be done before or after capping the vial. The techniques have the same
advantages and disadvantages as mentioned previously.
Figure 3-5: Loading a Vial
Introduction • Page 3-6
Summary of Contents for 15-0800-074
Page 1: ...Versa User Manual www teledynetekmar com Versa Manual Part 15 0800 074 ...
Page 8: ...Versa User Manual www teledynetekmar com Chapter 1 Introduction ...
Page 21: ...Versa User Manual www teledynetekmar com Chapter 2 Installation Setup ...
Page 36: ...Versa User Manual www teledynetekmar com Chapter 3 Basic Operation ...
Page 44: ...Versa User Manual www teledynetekmar com Chapter 4 Versa TekLink 2G ...
Page 61: ...Versa User Manual www teledynetekmar com Chapter 5 21 CFR Part 11 Compliance ...
Page 77: ...Versa User Manual www teledynetekmar com Chapter 7 Maintenance Troubleshooting ...
Page 90: ...Versa User Manual www teledynetekmar com Chapter 8 Diagrams ...
Page 91: ...Versa User Manual www teledynetekmar com Detailed Plumbing Diagrams Page 8 2 ...
Page 100: ...Versa User Manual www teledynetekmar com Chapter 9 Index ...