MULTITOX DETECTOR
OPERATING MANUAL
44
NOSP17668
Revision 03
a
Contacts are then described as below:
Detector status
Relay 1
“Fault”
Relay 2
“Alarm”
Relay 3
“Alarm”
Normal
(no alarm, no fault, no inhibition
and detector powered)
C1-T1 closed
C1-R1 open
C2-T2 open
C2-R2 closed
C3-T3 open
C3-R3 closed
Alarm
C1-T1 closed
C1-R1 open
C2-T2 closed
C2-R2 open
C3-T3 closed
C3-R3 open
Ambiguity function
(TX explosimeter only)
C1-T1 open
C1-R1 closed
C2-T2 closed
C2-R2 open
C3-T3 closed
C3-R3 open
Fault or inhibition
C1-T1 open
C1-R1 closed
C2-T2 open
C2-R2 closed
C3-T3 open
C3-R3 closed
Maintenance Inhibition
(during maintenance mode)
Depend of configuration. See chapter 6.2
Power loss
C1-T1 open
C1-R1 closed
C2-T2 open
C2-R2 closed
C3-T3 open
C3-R3 closed
6.8.6.2
Output current configuration
This menu allows you to switch the format of the output current between 4-20 mA and 0-22 mA
.
Factory setting: the output current is 4-20 mA
6.8.7
[INH]ibition submenu
Maintenance inhibition can be configured in « frozen » mode (factory setting) or « free » mode.
•
In « frozen» mode, the outputs (current and relay) remain in their previous state.
•
For example, if the device displays a fault (2.0 mA), it will remain in this mode during the
inhibition.
•
If the device is configured in « free » mode, the current output will remain at the same level
than the permanent inhibition
Summary of Contents for Everywhereyoulook DG-TT7-S
Page 32: ...MULTITOX DETECTOR OPERATING MANUAL 26 NOSP17668 Revision 03a ...
Page 34: ...MULTITOX DETECTOR OPERATING MANUAL 28 NOSP17668 Revision 03a ...
Page 47: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 41 ...
Page 54: ...MULTITOX DETECTOR OPERATING MANUAL 48 NOSP17668 Revision 03a ...
Page 58: ...MULTITOX DETECTOR OPERATING MANUAL 52 NOSP17668 Revision 03a ...
Page 60: ...MULTITOX DETECTOR OPERATING MANUAL 54 NOSP17668 Revision 03a ...
Page 69: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 63 10 Declaration of conformity ...
Page 70: ...MULTITOX DETECTOR OPERATING MANUAL 64 NOSP17668 Revision 03a ...
Page 71: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 65 ...