6.8.2.2
Calibration
Calibration must be made with the gas the detector is set to detect, with the SET menu (F2 key) of
the wireless communication tool TLU600 and a calibration kit.
The calibration gas should be injected at a flow rate between 30 l/h and 60 l/h. For the DGi-
TT7-O version, on a 25%vol scale, ambient air at 21%vol O
2
generally fits.
For DG-TT7-S series, it is necessary to have a humidifier tube into the gas circuit. Use the specific
calibration kit.
Open the flow regulator (0.5 l / min), wait for stabilization (about 2-3 min)
and calibrate.
If one wants to carry out another bump test or calibration and expect an accurate measurement,
a 60 minutes clean air recovery is considered as a minimum interval. This interval increases up to
90 minutes if the gas exposure has been about 5 minutes and extended to, at least, 2 hours if the
gas exposure has exceeded 5 min.
NB : The humidifier tube provided in the calibration kit has to be change when the color becomes
dark brown
For DG-TT7-S versions, the H
2
S mixture must be in air and not in nitrogen
otherwise a significant calibration error is made
The detector is in maintenance inhibition mode for 10 minutes after it goes back to main screen.
Use the setting menu and validate the INH command for acknowledgement.
To end of communication between the TLU6000 and the detector is done by pressing F4 key on
main menu.
5
When the splash guard (AS019) is used with the calibration cup, stabilization time increases
slightly (approximately 1 minute) during an injection with cup.
Summary of Contents for Everywhereyoulook DG-TT7-S
Page 32: ...MULTITOX DETECTOR OPERATING MANUAL 26 NOSP17668 Revision 03a ...
Page 34: ...MULTITOX DETECTOR OPERATING MANUAL 28 NOSP17668 Revision 03a ...
Page 47: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 41 ...
Page 54: ...MULTITOX DETECTOR OPERATING MANUAL 48 NOSP17668 Revision 03a ...
Page 58: ...MULTITOX DETECTOR OPERATING MANUAL 52 NOSP17668 Revision 03a ...
Page 60: ...MULTITOX DETECTOR OPERATING MANUAL 54 NOSP17668 Revision 03a ...
Page 69: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 63 10 Declaration of conformity ...
Page 70: ...MULTITOX DETECTOR OPERATING MANUAL 64 NOSP17668 Revision 03a ...
Page 71: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 65 ...