WaveAce Remote Control
ACQUISITION - OFFSET, OFST
The OFFSET command allows adjustment of the vertical offset of the specified input channel. The
OFFSET? query returns the DC offset value of the specified channel.
The maximum ranges depend on the fixed sensitivity setting. If an out-of-range value is entered, the
oscilloscope is set to the closest possible value.
NOTE: The probe attenuation factor is not taken into account in offset adjustments. Also, the unit V is optional.
Command Syntax
<channel>:OFfSeT <offset>
<channel>:= {C1, C2, C3, C4}
<offset>:=
Refer to datasheet specification for your Teledyne LeCroy oscilloscope at
teledynelecroy.com.
Query Syntax
<channel>:OFfSeT?
Response Format
<channel>:OFfSeT <offset>
Example
The following instruction sets the offset of Channel 2 to -3 V:
C2:OFST -3V
24
920836 Rev B