o
While the device is sniffing click in the
Classic Encryption
link key/pin code field. This action
places the focus on that window.
o
Change the
/pin code.
o
The Status window at the bottom of the page will inform the user to move focus away from the
link key/pin code window.
o
Click the mouse outside the link key/pin code field or press the Tab key. This action will remove
the focus from the link key/pin code window.
o
The link key/pin
code changes are
automatically saved
to the configuration file.
4.1.2.1 BPA 600 Capture with ProbeSync
ProbeSync™ allows multiple ComProbe analyzers to work seamlessly together and to share a common clock.
Clock sharing allows the analyzers to precisely synchronize communications stream and to display resulting
packets in a single shared view.
If two ComProbe BPA 600 hardware are connected in a ProbeSync configuration, two to four links can be
synchronized. Four links result when each BPA 600 analyzer is configured for Classic Only Multiple
Connections with two links per BPA 600 device.
When configured for synchronization through ProbeSync one BPA 600 device provides the clock to the other
device. The clock is provided by a CAT 5 cable between the master BPA 600
OUT
connector—sending the
synchronizing clock—to the BPA 600 hardware
IN
connector—receiving the clock.
When the BPA 600 software runs in ProbeSync one
Control
window opens with two
BPA 600 datasource
windows, one for each connected device. Each device datasource is setup individually to sniff their respective
link. Should the hardware be connected incorrectly, that is
IN
to
IN
or
OUT
to
OUT
, an error message will
appear. Follow the instructions in error message. To continue click on the
OK
button. The
BPA 600
datasource Status
window will also display a warning message suggesting information sources.
Figure 4.5 - Incorrect ProbeSync Hardware Connection Error
TELEDYNE LECROY
Chapter 4 Capturing and Analyzing Data
67
Frontline BPA 600 Hardware & Software User Manual
Summary of Contents for Frontline BPA 600
Page 1: ...Hardware and Software User Manual Revision Date 3 14 2017...
Page 304: ...TELEDYNE LECROY Appendicies 296 Frontline BPA 600 Hardware Software User Manual...
Page 310: ...TELEDYNE LECROY Appendicies 302 Frontline BPA 600 Hardware Software User Manual...
Page 318: ...TELEDYNE LECROY Appendicies 310 Frontline BPA 600 Hardware Software User Manual...
Page 328: ...TELEDYNE LECROY Appendicies 320 Frontline BPA 600 Hardware Software User Manual...
Page 340: ......
Page 348: ...340 Frontline BPA 600 Hardware Software User Manual Appendicies...