Syringe Pump Module Installation Guide
Chapter 2: Installing the Syringe Module
8
3
Check that the black pin is in the position toward the front of the autosampler.
If necessary, turn the block over and move the pin to the other hole.
F
IGURE
2-2
Position of Pin on Mounting Block (Bottom View).
4
Use four hex-head screws to attach the mounting block to the frame of the
autosampler.
F
IGURE
2-3
Attaching the Mounting Block.
The installed mounting block should look like this:
F
IGURE
2-4
Inside Mounting Block After Installation.
5
If more than one module is needed on one side of the autosampler, then an
outside mounting block needs to be fitted to accommodate that module. The
outside mounting block can be fitted by positioning the white mounting pins
on the side of the block (as shown in
F
IGURE
2-1
) into the spaces in the side of
the inside mounting block (as can be seen in
F
IGURE
2-4
) and pushing the
outside mounting block down until it is flush with the inside mounting block.
Front
Back