1.1  General Description 

 

All systems comprise of the following components :- 
 

1 – SM6 Airflow Sensor, 
1 – AFA1000  Alarm unit, 
1 – DC power supply 
 
If the Sash Alarm System option is included there will also be a sash micro switch or proximity 
switch. 
 

Operator Features   

--- the alarm has the following operator features :- 

 

Digital Display  

 
The digital display is a back-lit, full graphic unit with a visual display of approx 56 x 27 mm. 
The display operates through the software allowing the generation of figures, wording and 
Icons. 
 
The display shows  the fume cupboard face velocity in 

m/sec

 or

 fpm 

when enabled or the 

alternative with no velocity reading but showing 

AIR FAIL / AIR SAFE

 as continuous display. 

All of the above are configurable via the alarm key pad. 
 
An ‘ 

event time line

 ’ segmented into 20 x 3 minute segments will scroll across the display  

( when enabled) .This takes the form of a graphical ‘ blip’ that will progress from the right hand 
side to the left hand side – representing events that have occurred during the past hour.  
The alternative to the event time line is a dynamic  ‘ 

bar graph

 ’ representing the face velocity 

 
The display shows a 

Horn

 icon ( with line through it ) when the audible alarm is in the Muted 

condition 
 

Sash High

 –  will be displayed when the Sash alarm is enabled and the sash is raised above 

 

 

the max safe working opening.  

 

 

This display will alternate on/off with the velocity reading. 

 

Ext Alarm

 –   will be displayed when the external alarm input is activated ( when enabled ) 

 

 

This display will alternate on/off with the velocity reading 

 

Air Fail  - 

will be displayed if the airflow is less than the Low air alarm point. 

 

 

This display will alternate on/off with the velocity reading  

 

High Air - 

will be displayed if the airflow is more than the High air alarm point. 

 

 

This display will alternate on/off with the velocity reading 

 

Set-back

  -    will be displayed if the night set-back function is activated ( when enabled ) 

 

 

This display will alternate on/off with the velocity reading 

 

Disabled

  -   will be displayed if the alarm disable function is activated ( when enabled ) 

 

 

This display will alternate on/off with the velocity reading 

 
 
 
 
 
 
 

 

Summary of Contents for AFA1000 / 1 - Mk2

Page 1: ...Digital display 3 Relay inputs 3 Relay outputs Com port Used for alarm indication and monitoring on Fume Cupboards Model AFA1000 2 Mk2 Digital display 3 Relay inputs 3 Relay outputs 3 Pushbuttons Com...

Page 2: ...Passwords in the Main Menu or alternatively use a Laptop connected to the Com port and the Upload Download software provided Velocity Bar Graph or Alarm Time Line Velocity display m sec or fpm Control...

Page 3: ...R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl Sash Micro switch REF...

Page 4: ...representing events that have occurred during the past hour The alternative to the event time line is a dynamic bar graph representing the face velocity The display shows a Horn icon with line through...

Page 5: ...the calibration and configuration menu or to select options or values External Connections the alarm unit will have the following connection points Input 1 volt free relay input configurable for norma...

Page 6: ...a Fan stop start button for the fume cupboard extract fan or as Night Set back Override Pushbutton 2 This button can be configured to act as a Lights ON OFF or UV Lights On OFF for the fume cupboard o...

Page 7: ...e configuration of the alarm NOTE If you enter the Calibration or Configure Menu by accident press the buttons at the same time to escape back to the Main Menu The menus and sub menus are in plain lan...

Page 8: ...ime the Low air alarm resets automatically High airflow If configured High Air toggles on off with display Audible alarm sounds can be muted via Enter pushbutton Audible alarm Mute When the audible al...

Page 9: ...ctivated Night Set Back Icon is displayed Red LED on Flashing Reduced Low air alarm if configured Audible alarm muted Mute Icon shown on display External alarm When input configured as External alarm...

Page 10: ...isables Night setback operation in Override On condition Note In the On condition the unit ignores any Night Set back input signals and remains in this condition until Off is selected Pushbutton 2 Pum...

Page 11: ...setup press ENTER to continue and in the Main Menu use the buttons on the alarm facia select SETUP and then press the ENTER button 7 In the Setup Menu select CALIBRATION and press the ENTER button 8...

Page 12: ...individual lowest point on the grid For most fume cupboards this low point is on the bottom row in the centre and is a convenient position to measure and for future reference when checking the alarm...

Page 13: ...ensor Hole for Vent Tube SM6 Sensor Dimensions Rear view SM6 Sensor Panel Cutout Dimensions Front view NOT TO SCALE 2 x Holes for self tapping fixing screw 4mm Fixing Hole provided on SM6 sensor Panel...

Page 14: ...ng position using a flexible tube With a single skin side wall it is possible to fix the sensor on the outside of the fume cupboard and connect directly to the inner chamber in the ideal position This...

Page 15: ...SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl...

Page 16: ...oximity switch Optional SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 I...

Page 17: ...F all other warranties and is subject to the conditions and LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE THE EXCLUSIVE RE...

Reviews: