WFM6100, WFM7000, and WFM7100 With Option FP Service Manual
4- 25
Removal and Replacement Procedures
This section contains information about removal and replacement of all modules
in the waveform monitor.
Preparation
WARNING.
Before doing this or any other procedure in this manual, read the
safety summaries found at the beginning of this manual. Also, to prevent possible
injury to service personnel or damage to the instrument components, read
Installation
in Section 2, and
Preventing ESD
in this section.
This subsection contains the following items:
H
This preparatory information that you need to properly do the procedures
that follow.
H
List of tools required to remove and disassemble all modules.
H
Procedures for removal and reinstallation of the modules.
WARNING.
Before doing any procedure in this subsection, disconnect the power
cord from the line voltage source. Failure to do so could cause serious injury or
death.
NOTE
. Read
Equipment Required
for a list of the tools needed to remove and
install modules in this instrument. See Table 4--9, on page 4--26.
Read the cleaning procedure before disassembling the instrument for cleaning.
Equipment Required.
Most modules in the instrument can be removed with a
screwdriver handle mounted with a size T-10, Torx
R
screwdriver tip. All
equipment required to remove and reinstall the modules is listed in Table 4--9.