WFM6100, WFM7000, and WFM7100 With Option FP Service Manual
3- 1
Adjustments
The waveform monitor does not normally require any adjustments. The
adjustments which can be made are semi-automated, and should only be made
while performing a Performance Verification. Make the appropriate adjustment if
the instrument fails one of these tests in the Performance Verification procedure:
H
HD Cable Meter
(WVR7100 Option EYE or PHY only)
H
SD Cable Meter
(WVR6100 Option EYE and WVR7100 Options SD and
EYE or PHY only)
H
HD Jitter Noise Floor
(WVR7100 Option EYE or PHY only)
H
Eye Gain
(WVR6100 and WVR7100 Option EYE or PHY only)
H
PixMon Output Level
(RGB and Option CPS)
H
Composite Analog Input DC Offset Restore Off
(Option CPS only)
H
Composite Analog Vertical Measurement Accuracy
(Option CPS only)
H
Composite Analog Frequency Response
(Option CPS only)
H
Analog Audio Level Meter Accuracy Over Frequency
(Options DA, AD, DD,
and DDE only)
After making the the adjustment shown here, repeat the failed Performance
Verification step.
Required Equipment
These adjustment procedures require the same equipment as the Performance
Verification procedure. See
Specifications and Performance Verification
for full
details.