Troubleshooting
6–52
TLA 704 Color Portable Mainframe Service Manual
Table 6–6: BIOS error codes and explanations (Cont.)
Error code
Action to be taken
Explanation
Parity Check 1
A parity error was found on the communications bus. The
BIOS will return the address of the error or display ???? if it
cannot.
Replace the controller board.
Parity Check 2
A parity error was found on the I/O bus. The BIOS will return
the address of the error, or display ???? if it cannot.
Replace the controller board.
Press <F1> to resume,
<F2> to Setup
This message is displayed after a recoverable error
message. Press the F1 key to restart the boot process, or
press the F2 key to enter the CMOS setup and make
changes to the definitions.
This is an informational message. Press the
F1 key to restart the boot process, or press the
F2 key to enter the CMOS setup and make
changes to the definitions. Verify that the CMOS
setup is configured correctly.
Press and hold CTRL-
ALT-F2 to enter SET-
UP
This is an optional message that can be turned on and off in
CMOS setup. It is only displayed during a power-on self-test.
This is an informational message.
Previous boot incom-
plete - Default configu-
ration used
The previous power-on self- test was not successful. If the
power-on self-test fails, it loads default values and displays a
message to run the CMOS setup. If incorrect values are
loaded, future boot attempts may fail.
This is an informational message. Verify that the
CMOS setup is configured correctly.
Real time clock error
The real time clock may be defective.
Replace the controller board.
Shadow Ram Failed at
offset: nnnn
The shadow RAM has failed at HEX location nnnn.
The system will run without shadow RAM, but it
will run slowly. Exchange the SIMMs; if one of
the SIMMs is defective, the error address
should change.
System battery is dead
- Replace and run
SETUP
The CMOS clock may be dead or disconnected. The CMOS
setup may have to be reset.
The system battery is internal to the RAM/Cal-
endar component. To replace the system
battery, replace the RAM/Calendar component.
Note:
After replacing the RAM/Calendar component,
you must reflash the BIOS. Refer to Update the
BIOS on page 6–53 and perform the procedures
to reflash the BIOS.
System BIOS shad-
owed
The system BIOS was copied to shadow RAM.
This is an informational message. No action
needs to be performed.
System cache error -
Cache disabled
When the system BIOS failed the cache test, the cache was
disabled.
The system will run without cache RAM, but it
will run slowly. Try rebooting the system.
System CMOS check-
sum bad - run SETUP
The system CMOS settings may be incorrect or modified
incorrectly by some application. The CMOS setup may have
to be re-defined.
This is an informational message. Reboot the
system, and then verify that the CMOS setup is
configured correctly.
System RAM Failed at
offset: nnnn
The system RAM at HEX address nnnn may be defective.
Exchange the SIMMs; if one of the SIMMs is
defective the error address should change.
System timer error
The BIOS failed the timer test. The controller board may be
defective.
Try rebooting the system first. If this message
keeps reappearing, replace the controller board.
Summary of Contents for TLA 704
Page 13: ...Service Safety viii TLA 704 Color Portable Mainframe Service Manual...
Page 17: ...Preface xii TLA 704 Color Portable Mainframe Service Manual...
Page 21: ...Introduction xvi TLA 704 Color Portable Mainframe Service Manual...
Page 22: ......
Page 35: ......
Page 44: ......
Page 48: ...Theory of Operation 3 4 TLA 704 Color Portable Mainframe Service Manual...
Page 49: ......
Page 51: ...Performance Verification 4 2 TLA 704 Color Portable Mainframe Service Manual...
Page 52: ......
Page 54: ...Adjustment Procedures 5 2 TLA 704 Color Portable Mainframe Service Manual...
Page 55: ......
Page 59: ...Maintenance 6 4 TLA 704 Color Portable Mainframe Service Manual...
Page 93: ...Removal and Installation Procedures 6 38 TLA 704 Color Portable Mainframe Service Manual...
Page 113: ...Repackaging Instructions 6 58 TLA 704 Color Portable Mainframe Service Manual...
Page 114: ......
Page 117: ......
Page 119: ...Electrical Parts List 8 2 TLA 704 Color Portable Mainframe Service Manual...
Page 120: ......
Page 123: ......
Page 141: ...Replaceable Parts List 10 18 TLA 704 Color Portable Mainframe Service Manual...
Page 142: ......
Page 143: ......