Performance Tests
TDS7104 & TDS7054 Service Manual
4- 29
Table 4- 4: Test equipment (Cont.)
Item number and
description
Purpose
Example
Minimum requirements
16. Adapter
Female N-to-male BNC
Tektronix part number
103-0058-00
Checking Analog Bandwidth
17. Adapter (three required)
SMA female-to-female
Tektronix part number
015-1012-00
Checking the delay between
channels
18. Adapter (three required)
SMA male-to-female BNC
Tektronix part number
015-1018-00
Checking the delay between
channels
19. Pulse Generator
2 MHz,
≤
150 ps rise time, 5 V out
Fluke 9500B
1,2
Used to Test Delta Time
Measurement Accuracy
20. Cable, Coaxial
(two required)
50
Ω
, 20 in, male-to-male SMA
connectors
Tektronix part number
174-1427-00
Used to Test Delta Time
Measurement Accuracy
21. Adapter
SMA “T”, male to 2 SMA female
Tektronix part number
015-1016-00
Used to Test Delta Time
Measurement Accuracy
22. Adapter
SMA female to BNC male
Tektronix part number
015-0572-00
Used to Test Delta Time
Measurement Accuracy
23. Adapter
BNC male to female elbow
Tektronix part number
103-0031-00
Used to Test Delta Time
Measurement Accuracy
24. Terminator
Short circuit, SMA connector
Tektronix part number
015-1021-00
Used to Test Delta Time
Measurement Accuracy
25. Attenuator, 2X
Ratio: 2X; impedance 50
Ω
; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0069-02
Used to Test Delta Time
Measurement Accuracy
1
Fluke 9500B/1100, 9500B/3200, or 9500B/2200 and an output head (9510, 9530, or 9560) appropriate for the bandwidth of
the oscilloscope being tested.
2
For Delta Time Measurement Accuracy, use a Fluke 9500B or a pulse generator with a rise time as shown in Table 4- 9 on
page 4- 64.
3
P6139A probe is not appropriate for the TDS7104 oscilloscopes.
4
Warning: The P6243 and P6245 probes that may be used with this oscilloscope provide an extremely low loading
capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure
signals exceeding
±
8 V, or errors in signal measurement will be observed. Above 40 V, damage to the probe may result.
To make measurements beyond
±
8 V, use either the P6139A probe (good to 500 V), or refer to the catalog for a recom-
mended probe. P6139A is not an appropriate probe for the TDS7104 oscilloscopes.
Summary of Contents for TDS7054
Page 4: ......
Page 14: ...Table of Contents x TDS7104 TDS7054 Service Manual...
Page 18: ...Service Safety Summary xiv TDS7104 TDS7054 Service Manual...
Page 22: ...Preface xviii TDS7104 TDS7054 Service Manual...
Page 56: ...Operating Information 2 12 TDS7104 TDS7054 Service Manual...
Page 60: ...Theory of Operation 3 4 TDS7104 TDS7054 Service Manual...
Page 64: ...Performance Verification 4 4 TDS7104 TDS7054 Service Manual...
Page 76: ...Brief Procedures 4 16 TDS7104 TDS7054 Service Manual...
Page 86: ...Performance Tests Semi Automated Method 4 26 TDS7104 TDS7054 Service Manual...
Page 144: ...Performance Tests 4 84 TDS7104 TDS7054 Service Manual...
Page 160: ...Maintenance 6 6 TDS7104 TDS7054 Service Manual...
Page 230: ...Troubleshooting 6 76 TDS7104 TDS7054 Service Manual...
Page 232: ...Repackaging Instructions 6 78 TDS7104 TDS7054 Service Manual...
Page 238: ...Electrical Parts List 8 2 TDS7104 TDS7054 Service Manual...