Performance Tests
4- 66
TDS7104 & TDS7054 Service Manual
attenuator. See Figure 4--29. The 10X attenuator is optional if the SG503
is used.
TDS7000 oscilloscope
50
Ω
coaxial cable
Sine wave generator
Output
10X attenuator
Figure 4- 29: Initial test hookup
2.
Confirm the trigger system is within time-accuracy limits for pulse-glitch or
pulse-width triggering (time range
≤
500 n
s)
:
a.
Display the test signal:
Set the output of the sine wave generator for a
100 MHz, five-division sine wave on screen. Press
PUSH TO
SET 50%
.
b.
Set the trigger mode:
Press the Trigger
MODE
button to toggle it to
NORMAL
.
c.
Set upper and lower limits that ensure triggering:
See Figure 4--30.
H
Press the front-panel
ADVANCED
button and select the
A Event
tab; then select width triggering by touching the
Width
button.
Summary of Contents for TDS7054
Page 4: ......
Page 14: ...Table of Contents x TDS7104 TDS7054 Service Manual...
Page 18: ...Service Safety Summary xiv TDS7104 TDS7054 Service Manual...
Page 22: ...Preface xviii TDS7104 TDS7054 Service Manual...
Page 56: ...Operating Information 2 12 TDS7104 TDS7054 Service Manual...
Page 60: ...Theory of Operation 3 4 TDS7104 TDS7054 Service Manual...
Page 64: ...Performance Verification 4 4 TDS7104 TDS7054 Service Manual...
Page 76: ...Brief Procedures 4 16 TDS7104 TDS7054 Service Manual...
Page 86: ...Performance Tests Semi Automated Method 4 26 TDS7104 TDS7054 Service Manual...
Page 144: ...Performance Tests 4 84 TDS7104 TDS7054 Service Manual...
Page 160: ...Maintenance 6 6 TDS7104 TDS7054 Service Manual...
Page 230: ...Troubleshooting 6 76 TDS7104 TDS7054 Service Manual...
Page 232: ...Repackaging Instructions 6 78 TDS7104 TDS7054 Service Manual...
Page 238: ...Electrical Parts List 8 2 TDS7104 TDS7054 Service Manual...