![Tektronix TDS7054 Service Manual Download Page 136](http://html1.mh-extra.com/html/tektronix/tds7054/tds7054_service-manual_1078193136.webp)
Performance Tests
4- 76
TDS7104 & TDS7054 Service Manual
Check output
Figure 4- 34: Measurement of main trigger out limits
3.
Confirm SIGNAL OUT is within limits for gain:
a.
Measure gain:
H
Move the precision 50
Ω
cable from the
AUX OUT
BNC to the
SIGNAL OUT
BNC.
H
Set Vertical
SCALE
to 100 mV.
H
Press
PUSH TO SET 50%
.
H
From the button bar, touch
MEAS
and select the
Ampl
tab.
H
Touch the
Pk-Pk
button.
H
Touch
Close
.
b.
Check against limits:
H
From the button bar, touch the
Vert
button. Touch the Termination
1 M
Ω
button.
H
CHECK that the readout
Ch2 Pk-Pk
is between 80 mV and 120 mV,
inclusive.
H
Enter the voltage on the test record.
H
From the button bar, touch the
Vert
button. Touch the Termination
50
Ω
button.
H
Touch
Close
.
Summary of Contents for TDS7054
Page 4: ......
Page 14: ...Table of Contents x TDS7104 TDS7054 Service Manual...
Page 18: ...Service Safety Summary xiv TDS7104 TDS7054 Service Manual...
Page 22: ...Preface xviii TDS7104 TDS7054 Service Manual...
Page 56: ...Operating Information 2 12 TDS7104 TDS7054 Service Manual...
Page 60: ...Theory of Operation 3 4 TDS7104 TDS7054 Service Manual...
Page 64: ...Performance Verification 4 4 TDS7104 TDS7054 Service Manual...
Page 76: ...Brief Procedures 4 16 TDS7104 TDS7054 Service Manual...
Page 86: ...Performance Tests Semi Automated Method 4 26 TDS7104 TDS7054 Service Manual...
Page 144: ...Performance Tests 4 84 TDS7104 TDS7054 Service Manual...
Page 160: ...Maintenance 6 6 TDS7104 TDS7054 Service Manual...
Page 230: ...Troubleshooting 6 76 TDS7104 TDS7054 Service Manual...
Page 232: ...Repackaging Instructions 6 78 TDS7104 TDS7054 Service Manual...
Page 238: ...Electrical Parts List 8 2 TDS7104 TDS7054 Service Manual...